Unlabelled: OBJECTVIE: Intraoperative neuromonitoring of thoracic-level pedicle screw implantation for detecting breaches in the pedicle cortex has adopted methods originally developed in the early 1990s for stainless steel (SS) alloy screws used at lumbosacral levels. In our recent attempts to monitor thoracic-level pedicle screw placement, we were surprised to find that these widely used stimulation parameters were largely ineffectual when stimulating directly through titanium alloy (Ti-alloy) pedicle screws. The objectives of this study, then, were twofold: (1) to report the number of episodes in which intraoperative neuromonitoring of thoracic screw position failed to detect a medially directed breach (or malplacement) in a previously described and limited sample set; and (2) to compare the frequency-specific impedance of a sample of Ti-alloy pedicle screws to comparably sized screws made of SS alloys. We predicted that Ti-alloy screws would demonstrate impairment in conduction properties that could help explain the difficulties we, and others, have recently experienced with neuromonitoring of thoracic pedicle screw placement.

Methods: Based on threshold values for train-of-four stimulation of spinal motor pathways, we quantified the incidence of medial breaches of thoracic-level pedicles in a small cohort of subjects. We also evaluated the conductive properties of Ti-alloy pedicle screws and compared these with SS screws. Eleven pedicle screws were examined using energy-dispersive x-ray spectroscopy to identify their alloys, after which DC resistance and AC impedance for each screw was measured. Furthermore, a subset of five screws was used to investigate the current delivery under dynamic testing conditions.

Results: Postoperative computed tomography of 6 subjects revealed 10 instances of significant medial screw malpositioning, out of a total of 88 screws placed. In each of these 10 instances, direct stimulation of thoracic pedicle screws at intensities considered in the literature to be clinically significant (i.e., ≤11 mA) failed to predict these medial pedicle breaches, yet each breach was reliably identified with low-intensity stimulation applied via a ball-tipped probe. For in vitro studies, most screws made of titanium alloys had higher resistance and impedance at tested frequencies compared with their SS counterparts. Moreover, there was widespread variability in conduction properties between Ti-alloy screws, whereas SS screws behaved in a more homogeneous manner.

Conclusions: When compared with screws made of SS, most Ti-alloy pedicle screws behaved more like semiconductors, showing conduction properties that were highly frequency dependent. These properties likely contributed to the difficulties we encountered in interpreting thoracic screw placements based on stimulus-evoked electromyography from direct screw stimulation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/WNP.0b013e3182767aacDOI Listing

Publication Analysis

Top Keywords

pedicle screws
28
screws
17
ti-alloy pedicle
16
intraoperative neuromonitoring
12
pedicle
12
pedicle screw
12
conduction properties
12
direct stimulation
8
thoracic-level pedicle
8
screw
8

Similar Publications

Preserving Cervical Mobility: A Novel Robot-Assisted Approach for Atlas Fracture Fixation.

Am J Case Rep

January 2025

Department of Orthopedic Surgery, Sichuan Provincial People's Hospital, School of Medicine, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.

BACKGROUND The management of unstable atlas fractures remains a subject of ongoing debate and controversy. The conservative surgical treatment commonly involves fusion, resulting in severe loss of cervical spine mobility, and a large incisions and extensive tissue dissection are required. We aim to introduce a novel concept and surgical approach for treating atlas fracture, one that involves minimizing trauma while maintaining mobility of the upper cervical spine without resorting to fusion.

View Article and Find Full Text PDF

Delayed diagnosis of spinal osteoblastoma presenting with radicular pain and scoliosis: A case report.

Int J Surg Case Rep

January 2025

University of Tunis El Manar, Faculty of Medicine of Tunis, 1007, Tunisia; Department of Orthopedic Surgery, Hospital Mongi Slim La Marsa, Tunisia.

Introduction And Importance: Osteoblastoma is a rare benign bone tumor, accounting for 1 % of primary bone tumors, often affecting the spine and sacrum. Accurate diagnosis is essential for appropriate treatment and prognosis.

Case Presentation: A 19-year-old male presented with two years of persistent nocturnal radicular and low back pain unresponsive to anti-inflammatory medications.

View Article and Find Full Text PDF

This descriptive longitudinal study aims to assess the risk factors for severe thoracic and lumbar vertebral compression fractures before and after surgery, contributing to preventive knowledge enhancement in communities and effective treatment management. The study involved 34 patients diagnosed with thoracic and lumbar vertebral compression fractures requiring surgery with bio-cement-augmented pedicle screws between June 2021 and June 2022. Postoperative complications, notably adjacent segment injury, were monitored, and patients received osteoporosis management post-surgery.

View Article and Find Full Text PDF

Cement-augmented pedicle screw instrumentation is a widely accepted method for managing osteoporotic fractures, but it carries inherent risks, particularly related to cement leakage and embolism. This study aimed to analyze a clinical case of complications following cement fixation and provide a detailed review of relevant literature. A 70-year-old patient underwent transpedicular screw instrumentation from L2-L4 with polymethyl methacrylate augmentation, which resulted in cement leakage into the spinal canal and subsequent pulmonary embolism.

View Article and Find Full Text PDF

Osteoporotic vertebral fractures (OVFs) in elderly patients pose challenges due to bone destruction and surgical risks. This case report describes a minimally invasive approach using calcium phosphate cement (CPC) vertebroplasty and short fusion with cement augmentation of pedicle screws (CAPS) in a 91-year-old woman with severe OVF. The patient underwent CPC vertebroplasty at L1 and CAPS fixation at T12-L2, followed by osteoporosis medication.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!