Automatic reading of ANCA-slides: evaluation of the AKLIDES system.

Clin Dev Immunol

Laboratory of Clinical Immunology, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX Maastricht, The Netherlands.

Published: August 2013

AI Article Synopsis

  • The study evaluates an automated system called AKLIDES for recognizing ANCA patterns in patients with ANCA-associated vasculitis (AAV) using neutrophils.
  • C-ANCA was correctly identified by routine microscopy in up to 97% of cases, while AKLIDES detected it in up to 95%, showing decent accuracy.
  • The results suggest that while AKLIDES is promising for diagnosing AAV, sensitivity on ethanol-fixed neutrophils needs enhancement for better effectiveness.

Article Abstract

The ANCA consensus prescribes screening by indirect immunofluorescence on neutrophils. We evaluated the first automated ANCA-pattern recognition system. C-ANCA (n = 39) and P-ANCA (n = 40) samples were selected from patients with ANCA-associated vasculitis (AAV). Non-AAV controls included sera from healthy controls (n = 40), sera with possible interfering antibodies (n = 46), or miscellaneous ANCA reactivity (n = 31). ANCA slides were analysed by AKLIDES and routine fluorescence microscopy. The C-ANCA pattern was recognized by routine microscopy in 92% and 97% on ethanol- and formalin-fixed slides, respectively. AKLIDES reported C-ANCA in 74% and 95%, respectively. P-ANCA was recognized by routine microscopy on ethanol-fixed neutrophils in 90%, while AKLIDES reported P-ANCA in 80%. Typically, only 65% and 33% of these samples showed the expected C-ANCA on formalin-fixed neutrophils by routine microscopy and AKLIDES, respectively. A C- or P-ANCA pattern was observed on ethanol-fixed neutrophils in 28% and 23% of the controls by routine microscopy and AKLIDES, respectively. Only 5% of the controls revealed C-ANCA on formalin-fixed neutrophils by routine microscopy and AKLIDES. Altogether, automated ANCA-pattern recognition by AKLIDES is promising. Distinction of C- and P-ANCA is good, but sensitivity on ethanol-fixed neutrophils needs improvement. When optimized, pattern recognition software may play an important role in AAV diagnostics.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3501831PMC
http://dx.doi.org/10.1155/2012/762874DOI Listing

Publication Analysis

Top Keywords

routine microscopy
20
ethanol-fixed neutrophils
12
microscopy aklides
12
aklides
8
automated anca-pattern
8
anca-pattern recognition
8
recognized routine
8
aklides reported
8
c-anca formalin-fixed
8
formalin-fixed neutrophils
8