Purpose: To develop a new velocity-selective (VS) excitation pulse sequence which is robust to field inhomogeneity, and demonstrate its application to non-contrast-enhanced peripheral MR angiography (MRA).

Methods: The off-resonance-robust VS saturation pulse is designed by incorporating 180° refocusing pulses into the k-space-based reference design and tailoring sequence parameters in a velocity region of interest. The VS saturation pulse is used as magnetization preparation for non-contrast-enhanced peripheral MRA to suppress background tissues but not arterial blood based on their velocities. Non-contrast-enhanced peripheral MRA using the proposed VS preparation was tested in healthy volunteers and a patient with arterial stenosis.

Results: Calf angiograms obtained using the new VS preparation show more uniform background suppression than the reference VS preparation, as demonstrated by larger mean values and smaller standard deviations of artery-to-vein and artery-to-muscle contrast-to-noise ratios (71.0 ± 11.4 and 75.3 ± 12.1 versus 61.7 ± 22.7 and 58.5 ± 27.8). Two-station peripheral MRA using the new VS preparation identifies stenosis of the femoral and popliteal arteries in the patient, as validated by digital subtraction angiography.

Conclusion: Non-contrast-enhanced MRA using the new VS magnetization preparation can reliably provide high angiographic contrast in the lower extremities with significantly improved immunity to field inhomogeneity.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3594489PMC
http://dx.doi.org/10.1002/mrm.24561DOI Listing

Publication Analysis

Top Keywords

non-contrast-enhanced peripheral
16
magnetization preparation
12
peripheral mra
12
preparation non-contrast-enhanced
8
peripheral angiography
8
field inhomogeneity
8
saturation pulse
8
preparation
7
non-contrast-enhanced
5
peripheral
5

Similar Publications

Velocity-selective (VS) magnetization preparation has shown great promise for non-contrast-enhanced (NCE) magnetic resonance angiography (MRA) with the ability to generate positive angiographic contrast directly using a single 3D acquisition. However, existing VS-MRA methods have an issue of aliased saturation around a certain velocity, known as velocity field-of-view (vFOV), which can cause undesired signal loss in arteries. This study aimed to develop a new version of the VS preparation pulse sequence that overcomes the aliased saturation problem in conventional VS preparation.

View Article and Find Full Text PDF

Evaluating anemia in HIV-infected patients using chest CT.

Open Med (Wars)

July 2024

Department of AIDS, Kunming Third People's Hospital/Yunnan Clinical Medical Center for Infectious Diseases, No. 319, Wujing Road, Guandu District, Kunming, 650041, Yunnan, China.

Objective: The aim of this study was to investigate the role of the Hounsfield unit value of chest CT non-contrast enhanced scan in evaluating the severity of anemia in HIV-infected patients.

Methods: Patients with HIV infection combined with anemia admitted to the Kunming Third People's Hospital were retrospectively collected and divided into mild anemia, moderate anemia, and severe anemia groups by peripheral hemoglobin (HB) content and calculated the ratio of ventricular septum density (VSD) to left ventricular density (LVD) and VSD to right ventricular density (RVD); then, the above patients were divided into the critical value group and the non-critical value group according to HB and compared the differences of LVD, RVD, VSD/LVD, and VSD/RVD in the two groups of patients.

Results: A total of 126 patients were included, with a mean age of 47.

View Article and Find Full Text PDF

Vein of Galen malformation (VGM) is a rare congenital, uncommon intracerebral vascular anomaly rarely complicated with the development of brain abscess as secondary to primary infection or after endovascular treatment. We report a very rare finding of a vein of Galen aneurysm associated with a large brain abscess at the time of diagnosis. A 12-year-old boy with a high-grade fever, severe headache, and recurrent episodes of convulsions came into the radiology department of Kassala Advanced Diagnostic Center.

View Article and Find Full Text PDF
Article Synopsis
  • - The study investigates the effectiveness of a new lower extremity MRA technique (TSE-Acc), which uses enhanced acceleration-selective arterial spin labeling, in diagnosing peripheral arterial disease (PAD) without relying on ECG or contrast agents, thus reducing risks like nephrogenic systemic fibrosis.
  • - Nine healthy volunteers and a patient with PAD underwent MRI scans to assess various metrics like signal intensity ratio (SIR), contrast-to-noise ratio (CNR), arterial visibility, and artifact contamination, using statistical tests for analysis with significance set at p < 0.05.
  • - The results indicated that while SIRs were similar across three MRA methods, TSE-Acc showed significantly higher CNRs in some arteries and
View Article and Find Full Text PDF

To develop a novel, unenhanced magnetic resonance angiography (MRA) exploiting cardiac-gated, single-slab 3D chemical-shift-encoded gradient- and spin-echo (GRASE) imaging for robust background suppression.The proposed single-slab 3D GRASE employs variable-flip-angles (VFA) in the refocusing radio-frequency (RF) pulse train to promote sensitivity to blood flow as well as imaging encoding efficiency. Phase encoding blips are inserted between adjacent lobes of the switching readout gradients such that chemical shift-induced phase information is encoded into different locations in k-space.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!