The hemodynamic changes that may occur in patients undergoing aortic balloon valvuloplasty were examined in the circulatory model. Four conclusions were reached. (1) Significant transvalvular pressure gradients appear only if the orifice is severely narrowed. (2) The magnitude of this gradient is highly flow dependent. (3) At critical narrowings, minute alterations in orifice size may induce most significant changes in the transvalvular gradient. (4) In low flow states significant gradients appear only if the stenosis is extreme. In patients with aortic stenosis, especially those with failing hearts and low cardiac output, the pressure gradient may be effectively decreased by minimal dilatation of the aortic orifice. These patients, however, remain in jeopardy because recurrent narrowing may cause a gradient incompatible with life.