Assessment of ADVIA Centaur analyzer for the measurement of 25-OH vitamin D.

Clin Lab

Servicio de Análisis Clínicos, Area Sanitaria de Ferrol, Avenida da residencia s/n, 15405 Ferrol, A Coruña, Spain.

Published: December 2012

Background: The aim of the present study was to evaluate some analytical performances of the ADVIA Centaur analyzer for the quantitative measurement of 25-OH Vitamin D [25(OH)D] in serum.

Methods: Serum concentrations of 25(OH)D were determined by a new automated chemiluminescence immunoassay method introduced by Siemens and adapted to an ADVIA Centaur analyzer, and compared with HPLC and a commercial chemiluminescence immunoassay (Liaison DiaSorin).

Results: The assay displayed a low intra-day (CV < 7.0%) and a low inter-day imprecision (CV < 9.0%). The ADVIA Centaur demonstrated a stronger Spearman's correlation (r = 0.921), better agreement (bias = -0.3 ng/mL), and better concordance correlation coefficient of Lin (P(c) = 0.88) and better kappa index (k = 0.92) with HPLC, than the Liaison DiaSorin assay (r = 0.907, bias = 5.9 ng/mL, p(c) = 0.84, k = 0.60) with HPLC. On the other hand, a significant inverse relationship was observed between parathyroid hormone (PTH) and 25(OH)D concentrations.

Conclusions: The ADVIA Centaur method is reliable for routine 25(OH)D determination in clinical laboratories

Download full-text PDF

Source

Publication Analysis

Top Keywords

advia centaur
20
centaur analyzer
12
measurement 25-oh
8
25-oh vitamin
8
chemiluminescence immunoassay
8
centaur
5
assessment advia
4
analyzer measurement
4
vitamin background
4
background aim
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!