Comparison of two portable lactate meters in dogs.

J Am Anim Hosp Assoc

Veterinary Specialty Services, Manchester, MO, USA.

Published: February 2013

The objective of this study was to compare two portable lactate analyzers for testing canine patients in a clinical setting with a previously accepted methodology for measuring plasma lactate. Between Jan 1, 2005 and Jun 1, 2006, all samples were analyzed using two different portable analyzers (devices A and B), and 86 of those samples were also analyzed by a reference laboratory (REF). The concordance correlation coefficient (CCC) for device A and the REF was 0.949 (95% confidence interval [95% CI], 0.923-0.966). For device B and REF, the CCC was 0.990 (95% CI, 0.985-0.993). Only 8 of 85 samples (9.4%) exceeded a relative difference of 20% for device B (note that 1 of the 86 samples was not included because the lactate level fell below the detectable limit of device B), but 48 of 85 samples (56.5%) exceeded a relative difference of 20% for device A. Both portable lactate analyzers appeared to be effective in detecting clinically significant elevations in plasma lactate compared with the REF but device B had better agreement. A positive correlation among all three analyzers existed; however, the analyzers do not yield directly interchangeable results.

Download full-text PDF

Source
http://dx.doi.org/10.5326/JAAHA-MS-5700DOI Listing

Publication Analysis

Top Keywords

portable lactate
12
lactate analyzers
8
plasma lactate
8
samples analyzed
8
exceeded relative
8
relative difference
8
difference 20%
8
20% device
8
lactate
6
device
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!