A DT neutron yield diagnostic based on the reactions, (63)Cu(n,2n)(62)Cu(β(+)) and (65)Cu(n,2n)( 64) Cu(β(+)), has been fielded at the National Ignition Facility (NIF). The induced copper activity is measured using a NaI γ-γ coincidence system. Uncertainties in the 14-MeV DT yield measurements are on the order of 7% to 8%. In addition to measuring yield, the ratio of activities induced in two, well-separated copper samples are used to measure the relative anisotropy of the fuel ρR to uncertainties as low as 5%.