Background: Selected reaction monitoring (SRM) is a reliable mass spectrometry (MS)-based technique for quantification of small molecules. However, it is not applicable to matrix-assisted laser desorption/ionization time-of-flight/time-of-flight tandem MS (MALDI-TOF/TOF MS) instruments. This work presents a novel comparable MALDI-TOF/TOF MS technique, "Parallel Fragmentation Monitoring" (PFM), for high-throughput quantification of citrulline.

Method: Calibrator/sample solutions were spiked with internal standard that was a stable isotopic analog with 1 mass unit heavier than citrulline. Both citrulline and internal standard were isolated and fragmented in parallel by MALDI-TOF/TOF MS in the presence of graphitized carbon nanoparticles as matrix. The ratio of the peak intensities of the selected fragment of citrulline to that of internal standard was used to calibrate/calculate the concentrations of citrulline in samples.

Results: Linear calibration curves were obtained in the range of 10-250 μmol/l citrulline with correlation coefficients ≥0.997. Stored calibration curve and batch-specific calibration curve produced highly similar measurement values. Within- and between-day CVs were 3.1-8.7% and 3.5-10.6%, respectively, illustrating the reliability and robustness of PFM.

Conclusion: Using citrulline for proof-of-concept, we have developed the PFM technique with tremendous potential for high-throughput quantification of amino acids and other small molecules.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2012.10.039DOI Listing

Publication Analysis

Top Keywords

internal standard
12
graphitized carbon
8
carbon nanoparticles
8
mass spectrometry
8
small molecules
8
high-throughput quantification
8
citrulline internal
8
calibration curve
8
citrulline
6
quantification
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!