Objectives: To determine the relationship between insertion torque and pullout strength of 3.5-mm-diameter cortical screws in cadaveric humeri with different bone mineral densities (BMDs).
Methods: Five pairs of human humeri from each of 3 BMD groups (normal, osteopenic, and osteoporotic) were used. Holes were drilled in each humerus, and maximum insertion torque (T(max)) was measured by tightening a screw until stripping occurred. In the remaining holes, screws were tightened to 50%, 70%, or 90% of the T(max). A servohydraulic testing machine pulled each screw out at 1 mm/s while resulting force and axial displacement were recorded at 10 Hz. The authors checked for an effect of insertion torque (percent T(max)) on pullout strength using a general linearized and latent mixed model (Stata10), controlling for cortical thickness and BMD (T-score).
Results: Pullout strength for normal and osteoporotic bone was greatest for screws inserted to 50% T(max) and was significantly greater than that at T(max) but not significantly different from that at 70% or 90% T(max). For osteopenic bone, pullout strength was greatest at 70% peak torque, but it was not significantly different from the pullout strength at the 50% or 90% T(max) levels.
Conclusions: Tightening screws beyond 50% T(max) does not increase pullout strength of the screw and may place bone at risk for damage that might result in loss of fixation. Even after adjusting for bone thickness and density, there is no clear relationship between pullout strength and screw torque.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/BOT.0b013e318279791f | DOI Listing |
ACS Appl Mater Interfaces
January 2025
School of Engineering, RMIT University, Bundoora, VIC 3083, Australia.
Graphene and its derivatives have been widely used as reinforcing nanofillers for high-performance polymer nanocomposites. The effectiveness of the reinforcement largely depends on the properties of the nanofiller-matrix interface, which can be represented by the interfacial shear strength (IFSS). This work systematically investigates IFSS enhancements for polyethylene (PE) nanocomposites reinforced by graphene origami (GOri) through molecular dynamics pull-out simulations.
View Article and Find Full Text PDFCureus
November 2024
Aerospace Engineering, Universiti Putra Malaysia, Kuala Lumpur, MYS.
Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques.
View Article and Find Full Text PDFJBJS Essent Surg Tech
December 2024
Department of Orthopaedics and Rehabilitation, Yale School of Medicine, New Haven, Connecticut.
Background: For complete disruption of the posterolateral corner (PLC) structures, operative treatment is most commonly advocated, as nonoperative treatment has higher rates of persistent lateral laxity and posttraumatic arthritis. Some studies have shown that acute direct repair results in revision rates upwards of 37% to 40% compared with 6% to 9% for initial reconstruction. In a recent study assessing the outcomes of acute repair of PLC avulsion injuries with 2 to 7 years of follow-up, patients with adequate tissue were shown to have a much lower failure rate than previously documented.
View Article and Find Full Text PDFPLoS One
December 2024
Bio-Inspired Technology Group, Faculty of Mechanical Engineering, Department of BioMechanical Engineering, Delft University of Technology, Delft, The Netherlands.
Pedicle screws have long been established as the gold standard for spinal bone fixation. However, their fixation strength can be compromised in cases of low bone density, particularly in osteoporotic bone, due to the reliance on a micro-shape lock between the screw thread and the surrounding bone. To address this challenge, we propose augmenting conventional pedicles screws with a curved compliant anchor.
View Article and Find Full Text PDFHeliyon
December 2024
Department of Civil, Environmental, and Construction Engineering, Texas Tech University, USA.
This study presents the design and experimental evaluation of advanced corrosion protection coatings for application on prestressing strands which are the core constituents of prestressed concrete structures such as bridges. Variety of self-heal coatings embodying corrective and protective phenomena in response to the degrading effects of corrosion have been designed and tested in simulated aggressive weathering conditions. Standard 7-wire prestressing strands coated with self-heal epoxy, self-heal toughened epoxy and hybrid epoxy coating systems were subjected to salt fog spray up to a duration of 2500 h, and 3M CalCl, 3M NaOH, saturated Ca(OH) solutions and distilled water up to 45 days duration.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!