Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams.

Phys Med

Center of Molecular Imaging, Radiotherapy and Oncology, Institut de recherche expérimentale et clinique, Université catholique de Louvain, Avenue Hippocrate 54, 1200 Brussels, Belgium; ICTEAM Institute, Université catholique de Louvain, Louvain-la-Neuve, Belgium. Electronic address:

Published: November 2013

Purpose: To evaluate the uncertainties and characteristics of radiochromic film-based dosimetry system using the EBT3 model Gafchromic(®) film in therapy photon, electron and proton beams.

Material And Methods: EBT3 films were read using an EPSON Expression 10000XL/PRO scanner. They were irradiated in five beams, an Elekta SL25 6 MV and 18 MV photon beam, an IBA 100 MeV 5 × 5 cm(2) proton beam delivered by pencil-beam scanning, a 60 MeV fixed proton beam and an Elekta SL25 6 MeV electron beam. Reference dosimetry was performed using a FC65-G chamber (Elekta beam), a PPC05 (IBA beam) and both Markus 1916 and PPC40 Roos ion-chambers (60 MeV proton beam). Calibration curves of the radiochromic film dosimetry system were acquired and compared within a dose range of 0.4-10 Gy. An uncertainty budget was estimated on films irradiated by Elekta SL25 by measuring intra-film and inter-film reproducibility and uniformity; scanner uniformity and reproducibility; room light and film reading delay influences.

Results: The global uncertainty on acquired optical densities was within 0.55% and could be reduced to 0.1% by placing films consistently at the center of the scanner. For all beam types, the calibration curves are within uncertainties of measured dose and optical densities. The total uncertainties on calibration curve due to film reading and fitting were within 1.5% for photon and proton beams. For electrons, the uncertainty was within 2% for dose superior to 0.8 Gy.

Conclusions: The low combined uncertainty observed and low beam and energy-dependence make EBT3 suitable for dosimetry in various applications.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2012.10.001DOI Listing

Publication Analysis

Top Keywords

elekta sl25
12
proton beam
12
beam
9
ebt3 films
8
therapy photon
8
photon electron
8
electron proton
8
proton beams
8
dosimetry system
8
calibration curves
8

Similar Publications

Phantom dosimetry and cancer risks estimation undergoing 6 MV photon beam by an Elekta SL-25 linac.

Appl Radiat Isot

September 2020

Department of Physics, Faculty of Sciences, Golestan University, Gorgan, 49138-15739, Iran. Electronic address:

The High-energy linear accelerator (linac) is a valuable tool and the most commonly used devices for external beam radiation treatments in subjects suffer from cancer. To estimate the dose deposited in several organs of a female patient due to pelvic irradiation by an Elekta SL-25 linac in 6 MV photon beam mode, the MCNPX code is used considering the most details of linac. The equivalent dose in different organs is computed according to the face down position (prone) of MIRD and UFRO phantoms.

View Article and Find Full Text PDF

The tissue- or water-equivalence of dosimetry phantoms used as substitutes for water is essential for absorbed dose measurements in radiotherapy. At our institution, a heterogeneous pelvic phantom that consists of stacked Nylon-12 layers has recently been manufactured for Gafchromic film dosimetry. However, data on the use of Nylon as tissue-mimicking media for dosimetric applications are scarce.

View Article and Find Full Text PDF

This study aims to reduce the uncertainty about the photoneutron dose produced over a course of radiotherapy with high-energy photon beams and evaluate photoneutron contamination-based secondary malignancy risk for different treatment modalities. Dosimetric measurements were taken in Philips SL25/75, Elekta Synergy Platform (Elekta AB, Stockholm, Sweden), Varian Clinac DHX High Performance systems (Varian Medical Systems, Palo Alto, CA), and Cyberknife Robotic Radiosurgery Unit (Accuray Inc., Sunnyvale, CA) using bubble detector for neutron dosimetry.

View Article and Find Full Text PDF

Monte Carlo (MC) based calculation methods for modeling photon and particle transport, have several potential applications in radiotherapy. An essential requirement for successful radiation therapy is that the discrepancies between dose distributions calculated at the treatment planning stage and those delivered to the patient are minimized. It is also essential to minimize the dose to radiosensitive and critical organs.

View Article and Find Full Text PDF

Purpose: We analysed the effects of field size, depth, beam modifier and beam type on the amount of in-field and out-of-field neutron contamination for medical linear accelerators (linacs).

Methods: Measurements were carried out for three high-energy medical linacs of Elekta Synergy Platform, Varian Clinac DHX High Performance and Philips SL25 using bubble detectors. The photo-neutron measurements were taken in the first two linacs with 18 MV nominal energy, whereas the electro-neutrons were measured in the three linacs with 9 MeV, 10 MeV, 15 MeV and 18 MeV.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!