Development of a biochemical oxygen demand sensor using gold-modified boron doped diamond electrodes.

Anal Chem

Department of Chemistry, Faculty of Mathematics and Science, University of Indonesia, Kampus UI Depok, Jakarta 16424, Indonesia.

Published: November 2012

Gold-modified boron doped diamond (BDD) electrodes were examined for the amperometric detection of oxygen as well as a detector for measuring biochemical oxygen demand (BOD) using Rhodotorula mucilaginosa UICC Y-181. An optimum potential of -0.5 V (vs Ag/AgCl) was applied, and the optimum waiting time was observed to be 20 min. A linear calibration curve for oxygen reduction was achieved with a sensitivity of 1.4 μA mg(-1) L oxygen. Furthermore, a linear calibration curve in the glucose concentration range of 0.1-0.5 mM (equivalent to 10-50 mg L(-1) BOD) was obtained with an estimated detection limit of 4 mg L(-1) BOD. Excellent reproducibility of the BOD sensor was shown with an RSD of 0.9%. Moreover, the BOD sensor showed good tolerance against the presence of copper ions up to a maximum concentration of 0.80 μM (equivalent to 50 ppb). The sensor was applied to BOD measurements of the water from a lake at the University of Indonesia in Jakarta, Indonesia, with results comparable to those made using a standard method for BOD measurement.

Download full-text PDF

Source
http://dx.doi.org/10.1021/ac302090yDOI Listing

Publication Analysis

Top Keywords

biochemical oxygen
8
oxygen demand
8
gold-modified boron
8
boron doped
8
doped diamond
8
linear calibration
8
calibration curve
8
l-1 bod
8
bod sensor
8
bod
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!