Split calibration curve: an approach to avoid repeat analysis of the samples exceeding ULOQ.

Bioanalysis

Department of DMPK & Toxicology, Sai Advantium Pharma Ltd, Building 1, Plot 2, Chrysalis Enclave, International Biotech Park, Phase 2, Hinjewadi, Pune, 411057, India.

Published: October 2012

Background: The current practice of using calibration curves with narrow concentration ranges during bioanalysis of new chemical entities has some limitations and is time consuming. In the present study we describe a split calibration curve approach, where sample dilution and repeat analysis can be avoided without compromising the quality and integrity of the data obtained.

Results: A split calibration curve approach is employed to determine the drug concentration in plasma samples with accuracy and precision over a wide dynamic range of approximately 0.6 to 15,000 ng/ml for dapsone and approximately 1 to 25,000 ng/ml for cyclophosphamide and glipizide. A wide dynamic range of concentrations for these three compounds was used in the current study to construct split calibration curves and was successfully validated for sample analysis in a single run.

Conclusion: Using this method, repeat analysis of samples can be avoided. This is useful for the bioanalysis of toxicokinetic studies with wide dose ranges and studies where the sample volume is limited.

Download full-text PDF

Source
http://dx.doi.org/10.4155/bio.12.219DOI Listing

Publication Analysis

Top Keywords

split calibration
16
calibration curve
12
curve approach
12
repeat analysis
12
analysis samples
8
calibration curves
8
wide dynamic
8
dynamic range
8
split
4
approach avoid
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!