Purpose: In this work, a passive tracking sequence employing a phase-only cross correlation (POCC) algorithm was studied with a focus on the in vivo applicability of the technique. Therefore, MR-guided needle interventions were performed in a phantom and two animal experiments.

Methods: The targeting accuracy was quantified in an agarose phantom with 15 fiducials. For each fiducial, the distance between needle trajectory and target point was measured. In a first animal experiment at 3 T, the prostate of a pig was punctured under POCC guidance. Second, POCC-based tracking was performed during a laser-induced thermal therapy procedure in peripheral porcine muscle tissue at 1.5 T.

Results: In the phantom experiment, the 15 fiducials were penetrated with a mean accuracy of 1.5 ± 0.9 mm (mean duration for one puncture about 2 min). In the first animal experiment, the center of the pig's right prostatic lobe was accurately punctured within 15 min. In the second, targeting and insertion of the needle could be performed within 5 min and a thermal lesion was successfully created.

Conclusion: Our initial experience with the POCC-based tracking sequence indicates that this technique has the potential as an accurate and versatile tool for in vivo MR-guided needle interventions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2012.09.002DOI Listing

Publication Analysis

Top Keywords

mr-guided needle
12
needle interventions
12
phase-only cross
8
cross correlation
8
correlation pocc
8
tracking sequence
8
animal experiment
8
pocc-based tracking
8
needle
5
passive marker
4

Similar Publications

Background: In MR-guided in-bore percutaneous needle interventions, typically 2D interactive real-time imaging is used for navigating the needle into the target. Misaligned 2D imaging planes can result in losing visibility of the needle in the 2D images, which impedes successful targeting. Necessary iterative manual slice adjustment can prolong interventional workflows.

View Article and Find Full Text PDF

A robotic needle implant device for MR-guided high-dose-rate (HDR) prostate brachytherapy was developed. This study aimed to assess the feasibility and spatial accuracy of HDR brachytherapy using the robotic device, for a single intraprostatic target point.Five patients were treated from November 2019-June 2022 with the robot.

View Article and Find Full Text PDF

Body-Mounted MR-Conditional Robot for Minimally Invasive Liver Intervention.

Ann Biomed Eng

August 2024

Institute of Robotics and Intelligent Machines, Georgia Institute of Technology, 801 Atlantic Dr NW, Atlanta, GA, 30332, USA.

Article Synopsis
  • MR-guided microwave ablation (MWA) is effective for treating small hepatocellular carcinoma (HCC) tumors, but manual needle insertion often leads to inaccuracies in the procedure.
  • This paper introduces a compact MR-conditional robot designed to enhance needle guidance, featuring stacked Cartesian XY stages with pneumatic actuation for precise control.
  • Initial experiments show a needle insertion accuracy of 2.6 ± 1.3 mm at an 80 mm depth, and future developments will aim to optimize the system and conduct animal trials for further validation.
View Article and Find Full Text PDF

MR-guided HDR prostate brachytherapy with teleoperated steerable needles.

J Robot Surg

October 2023

Department of Biomechanical Engineering, University of Twente, Enschede, The Netherlands.

Conformity of tumour volumes and dose plans in prostate brachytherapy (BT) can be constrained by unwanted needle deflections, needle access restrictions and visualisation limitations. This work validates the feasibility of teleoperated robotic control of an active steerable needle using magnetic resonance (MR) for guidance. With this system, perturbations can be counteracted and critical structures can be circumvented to access currently inaccessible areas.

View Article and Find Full Text PDF

MR-guided focal cryoablation of prostate cancer has often been selected as a minimally-invasive treatment option. Placing multiple cryo-needles accurately to form an ablation volume that adequately covers the target volume is crucial for better oncological/functional outcomes. This paper presents an MRI-compatible system combining a motorized tilting grid template with insertion depth sensing capabilities, enabling the physician to precisely place the cryo-needles into the desired location.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!