Background: The magnetic surgical camera is an emerging technology having the potential to improve visualization without taking up port site space. However, tilting the point of view downward/upward can be done only by constantly applying a pressure on the abdomen. This study aims to test the hypothesis that the novel concept of local magnetic actuation (LMA) is able to increase the tilt range available for a magnetic camera without the need for deforming the abdominal wall. The hypothesis that 2-port laparoscopic nephrectomy in fresh tissue human cadavers could be performed by using the LMA camera is also tested.

Methods: First, the 2 cameras were separately inserted, anchored, and moved inside the inflated abdomen. Tilting angles were quantified by image analysis while intra-abdominal pressure changes were monitored. Then, 5 two-port nephrectomies were performed by using the LMA camera while collecting quantitative outcomes.

Results: The magnetic camera required a constant pressure on the magnetic handle to achieve an average ±20° tilt from the horizontal position, with an average of 7 mm Hg loss of intra-abdominal pressure. The LMA camera allowed for 75° of tilt from the horizontal position with a resolution of ±1°, without any need to deform the abdomen. All the nephrectomies were completed successfully within an average time of 11 minutes.

Conclusion: LMA is an effective strategy to provide magnetic cameras with wide-range and high-resolution vertical motion without the need to deform the abdominal wall.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1553350612462458DOI Listing

Publication Analysis

Top Keywords

lma camera
12
local magnetic
8
magnetic actuation
8
human cadavers
8
magnetic camera
8
abdominal wall
8
performed lma
8
intra-abdominal pressure
8
tilt horizontal
8
horizontal position
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!