Purpose: To present a novel method for MR elastography (MRE) of the prostate at 3 Tesla using a modified endorectal imaging coil.
Materials And Methods: A commercial endorectal coil was modified to dynamically generate mechanical stress (contraction and dilation) in a prostate phantom with embedded phantom "lesions" (6 mm diameter) and in a porcine model. Resulting tissue displacements were measured with a motion-sensitive EPI sequence at actuation frequencies of 50-200 Hz. Maps of shear modulus G were calculated from the measured phase-difference shear-wave patterns.
Results: In the G maps of the phantom, "lesions" were easily discernible against the background. The average G values of regions of interest placed in the "lesion" (8.2 ± 1.9 kPa) were much higher than those in the background (3.6 ± 1.4 kPa) but systematically lower than values reported by the vendor (13.0 ± 1.0 and 6.7 ± 0.7 kPa, respectively). In the porcine model, shear waves could be generated and measured shear moduli were substantially different for muscle (7.1 ± 2.0 kPa), prostate (3.0 ± 1.4 kPa), and bulbourethral gland (5.6 ± 1.9 kPa).
Conclusion: An endorectal MRE concept is technically feasible. The presented technique will allow for simultaneous MRE and MRI acquisitions using a commercial base device with minor, MR-conditional modifications. The diagnostic value needs to be determined in further trials.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jmri.23850 | DOI Listing |
Radiol Adv
September 2024
Medical Biophysics, Western University, London, N6A 3K7, Canada.
Background: Sodium (Na) MRI of prostate cancer (PCa) is a novel but underdocumented technique conventionally acquired using an endorectal coil. These endorectal coils are associated with challenges (e.g.
View Article and Find Full Text PDFBJU Int
October 2024
Westmead Hospital, Westmead, New South Wales, Australia.
Objective: To determine the diagnostic test accuracy of multiparametric magnetic resonance imaging (mpMRI) in detecting seminal vesicle invasion (SVI).
Methods: The Medical Literature Analysis and Retrieval System Online (MEDLINE), PubMed, the Excerpta Medica dataBASE (EMBASE) and Cochrane databases were search up to May 2023. We included studies that investigated the accuracy of mpMRI in detecting SVI when compared to radical prostatectomy specimens as the reference standard.
Magn Reson Imaging
December 2024
Department of Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA 94143, USA; The Graduate Group in Bioengineering, University of California, San Francisco and Berkeley, Berkeley, CA 94720, USA; UCSF Helen Diller Family Comprehensive Cancer Center, University of California, San Francisco, San Francisco, CA 94143, USA. Electronic address:
Purpose: To establish the incidence, size, zonal location and Gleason Score(GS)/Gleason Grade Group(GG) of sparse versus dense prostate cancer (PCa) lesions and to identify the imaging characteristics of sparse versus dense cancers on multiparametric MRI (mpMRI).
Methods: Seventy-six men with untreated PCa were scanned prior to prostatectomy with endorectal-coil 3 T MRI including T2-weighted imaging, diffusion-weighted imaging and dynamic contrast-enhanced MRI. Cancerous regions were outlined and graded on the whole-mount, processed specimens, with tissue compositions estimated.
Asian J Surg
May 2024
Department of Radiology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital, Hangzhou Medical College), Hangzhou, Zhejiang, 310014, China. Electronic address:
Rectal cancer (RC) is the third most frequently diagnosed cancer worldwide, and the status of its circumferential resection margin (CRM) is of paramount significance for treatment strategies and prognosis. CRM involvement is defined as tumor touching or within 1 mm from the outermost part of tumor or outer border of the mesorectal or lymph node deposits to the resection margin. The incidence of involved CRM varied from 5.
View Article and Find Full Text PDFJ Comput Assist Tomogr
May 2024
Department of Urology, Oregon Health and Science University, Portland, OR.
We describe early ex vivo proof-of-concept testing of a novel system composed of a disposable endorectal coil and converging multichannel needle guide with a reusable clamp stand, embedded electronics, and baseplate to allow for endorectal magnetic resonance (MR) imaging and in-bore MRI-targeted biopsy of the prostate as a single integrated procedure. Using prostate phantoms imaged with standard T 2 -weighted sequences in a Siemens 3T Prisma MR scanner, we measured the signal-to-noise ratio in successive 1-cm distances from the novel coil and from a commercially available inflatable balloon coil and measured the lateral and longitudinal deviation of the tip of a deployed MR compatible needle from the intended target point. Signal-to-noise ratio obtained with the novel system was significantly better than the inflatable balloon coil at each of five 1-cm intervals, with a mean improvement of 78% ( P < 0.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!