Background: Compliant, self-adjusting compression technology is a novel approach for durable prosthetic fixation of the knee. However, the long-term survival of these constructs is unknown.
Questions/purposes: We therefore determined the survival of the Compress prosthesis (Biomet Inc, Warsaw, IN, USA) at 5 and 10 actuarial years and identified the failure modes for this form of prosthetic fixation.
Methods: We retrospectively reviewed clinical and radiographic records for all 82 patients who underwent Compress knee arthroplasty from 1998 to 2008, as well as one patient who received the device elsewhere but was followed at our institution. Prosthesis survivorship and modes of failure were determined. Followup was for a minimum of 12 months or until implant removal (median, 43 months; range, 6-131 months); 28 patients were followed for more than 5 years.
Results: We found a survivorship of 85% at 5 years and 80% at 10 years. Eight patients required prosthetic revision after interface failure due to aseptic loosening alone (n = 3) or aseptic loosening with periprosthetic fracture (n = 5). Additionally, five periprosthetic bone failures occurred that did not require revision: three patients had periprosthetic bone failure without fixation compromise and two exhibited irregular prosthetic osteointegration patterns with concomitant fracture due to mechanical insufficiency.
Conclusions: Compress prosthetic fixation after distal femoral tumor resection exhibits long-term survivorship. Implant failure was associated with patient nonadherence to the recommended weightbearing proscription or with bone necrosis and fracture. We conclude this is the most durable FDA-approved fixation method for distal femoral megaprostheses.
Level Of Evidence: Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3563794 | PMC |
http://dx.doi.org/10.1007/s11999-012-2635-6 | DOI Listing |
Prosthet Orthot Int
January 2025
Mechanical and Mechatronics Engineering Department, University of Waterloo, Waterloo, Ontario, Canada.
Knee bracing is commonly used for rehabilitation after ligament surgery. However, the effectiveness of knee bracing in preventing ligament injuries is not widely studied. This study aimed to develop a computational methodology to investigate the effectiveness of a novel type of cable-stabilized knee brace on anterior cruciate ligament (ACL) strain during single-leg jump landing.
View Article and Find Full Text PDFBMC Musculoskelet Disord
January 2025
Department of Orthopedics, Lishui Central Hospital, Zhejiang, 323000, China.
Background: Posterolateral tibial plateau fractures pose significant challenges for orthopedic surgeons due to the anatomical risks associated with the posterolateral approach. Despite numerous surgical techniques available, there lacks a consensus on the optimal approach.
Methods: Articular line incision approach was employed on 12 patients suffering from posterolateral tibial plateau fractures.
J Orthop Case Rep
January 2025
Department of Orthopaedics, SN Medical College, Agra, Uttar Pradesh, India.
Introduction: Posterior cruciate ligament (PCL) avulsion fractures of the tibia with ipsilateral tibial shaft represent a rare but challenging orthopedic injury, necessitating careful consideration of surgical interventions for optimal outcomes. This case report presents the successful management of tibial shaft fracture (proximal 1/3rd junction) along with ipsilateral PCL avulsion fracture of tibia using a novel approach using open reduction internal fixation (ORIF) of tibial shaft with Locking compression plate (LCP) with cannulated cancellous screw and spiked washer fixation of PCL avulsion. There are no case reports or research articles available for the management of PCL avulsion fracture of tibia associated with ipsilateral tibial shaft fracture.
View Article and Find Full Text PDFMagn Reson Imaging
January 2025
Department of Radiology, University Hospital of Strasbourg, 1 Avenue Moliere, 67098 Strasbourg, France.
Purpose: Compressed Sensing (CS) is an emerging technique to accelerate MRI acquisitions. The aim of this study was to assess the reliability and accuracy of cartilage thickness measurements in the knee using a CS-enabled isotropic 3D Fast Spin-Echo (FSE) sequence on a 3-T MRI scanner.
Methods: Twenty-eight tibial condyle sections were collected from 14 adult patients who underwent total knee arthroplasty.
Purpose Of The Study: Intraarticular fractures of the distal femur rank among the most severe musculoskeletal injuries. Various treatment options, such as plate osteosynthesis or retrograde nailing, can be employed. This study aims to evaluate the clinical outcomes and complications of intraarticular distal femoral fractures treated with retrograde femoral nail, with particular emphasis on C3 fractures.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!