Purpose: This study tests the hypothesis that the design changes incorporated in the newer generation Triathlon posterior-stabilised TKA design result in kinematics that more closely reproduce the kinematics observed in healthy knees than those achieved by the older generation Scorpio posterior-stabilised TKA design.
Methods: Eleven patients with Triathlon posterior-stabilised TKA, twelve patients with Scorpio posterior-stabilised TKA, and 22 subjects with normal asymptomatic knees underwent fluoroscopic assessment of the knee during a step-up exercise and a weight-bearing deep knee bend. Two-dimensional and three-dimensional knee kinematics were assessed including the maximum flexion, the patella tendon angle (PTA), the patella flexion angle (PFA), the minimum distance between cam and post, and the tibio-femoral contact positions.
Results: The average maximum flexion achieved was 114° (SD 3°), 91° (SD 10°), and 143° (SD 14°) for the Triathlon, Scorpio, and Normal groups. The average cam/post mechanism engagement was at 63° (SD 24°) and 82° (SD 16°) for the Triathlon and Scorpio groups. The condylar contact points showed a paradoxical anterior slide for the Scorpio group which was not present in the Triathlon group. The PTA and PFA values of both implants showed significant differences from normal.
Conclusion: Overall, the Triathlon implant design, as compared to Scorpio TKA, produced kinematics closer to that of normal knees as proposed by the hypothesis. However, despite being closer to normal, the kinematics exhibited by the Triathlon group were still different from normal. A comparison of kinematic performance, taking into account altered design parameters, will contribute to improved understanding and future design considerations.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00167-012-2233-x | DOI Listing |
Cureus
December 2024
Clinical Research and Medical Writing, Meril Life Sciences Private Limited, Vapi, IND.
Aim The primary objective of the study was to evaluate the mid-term implant survivability, rate of revisions, and clinical and functional outcomes following patella resurfacing during total knee arthroplasty (TKA) utilizing posterior stabilized (PS) total knee system (TKS). Methods A prospective, single-arm, multi-center, post-marketing surveillance encompassed patients with end-stage primary knee osteoarthritis (OA) or inflammatory arthritis. The time points of the study included baseline, six weeks, six months, one year, and three years post-operatively.
View Article and Find Full Text PDFSICOT J
January 2025
Department of Orthopaedic Surgery, Joint Replacement Unit, Kuala Lumpur Hospital, Ministry of Health Malaysia, Jalan Pahang, 50586 Kuala Lumpur, Malaysia.
Introduction: Extraarticular deformity (EAD) with knee arthritis is a complex condition involving tri-planar bone deformity with pathological malalignment and chronic soft tissue contracture or laxity in the knee joint. Intraarticular correction by TKA, which was previously performed with conventional manual jig by mechanical alignment technique, had its limits and difficulties especially extensive soft tissue release and risk of jeopardizing the collateral ligaments. Robotic technology allows for reproducible and precise execution of surgical plan and allows adjustment to various new personalised alignment philosophy including functional alignment (FA).
View Article and Find Full Text PDFBMC Musculoskelet Disord
January 2025
Division of Orthopedic Surgery, Changhua Christian Hospital, Changhua, Taiwan.
Background: Despite advancements in prosthetic designs and surgical techniques, patellar dislocation remains a rare but significant complication following total knee arthroplasty, with an incidence ranging between 0.15% and 0.5%.
View Article and Find Full Text PDFKnee Surg Sports Traumatol Arthrosc
December 2024
Sezione di Chirurgia Protesica ad Indirizzo Robotico, Unità di Traumatologia Dello Sport, UOC Ortopedia e Traumatologia, Brescia, Italy.
Purpose: The application of robotics in revision arthroplasty particularly from unicompartmental knee arthroplasty (UKA) to total knee arthroplasty (TKA), is underexplored. The purpose of this study is to describe the surgical technique of an imageless robotic system used in the revision of UKA to TKA and to evaluate short- to mid-term outcomes.
Methods: This prospective study includes 35 patients treated from May 2020 to July 2023.
Knee
December 2024
Department of Orthopaedic Surgery, Seoul National University College of Medicine, Seoul, South Korea; Department of Orthopaedic Surgery, Seoul National University Hospital, Seoul, South Korea. Electronic address:
Background: Total knee arthroplasty (TKA) mainly involves the femoral, tibial, and bearing implants. Bearings are categorized by posterior cruciate ligament (PCL) status (cruciate-retaining [CR] vs. posterior-stabilized [PS]) and motion (mobile vs.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!