Unlabelled: It was recently found that after storage of a live viral vaccine at -80 °C in glass vials closed with rubber stoppers, a phenomenon was revealed which had not been observed before with other viral products stored at -20 °C: overpressure in the vials. As this phenomenon poses a serious safety problem for medical personnel as well as for the product itself, an investigation was initiated to identify the root cause of the overpressure. After exclusion of possible root causes (differences in air temperature or atmospheric air pressure during filling and quality control testing, outgassing from the formulation buffer) the remaining hypothesis involved a possible container closure integrity issue at low temperature. The glass transition temperatures (T(g)) of many rubber stopper formulations are in the range -55 to -70 °C. At storage temperatures below T(g), the rubber stopper loses its elastic properties and there is a risk that the seal integrity of the vial could be compromised. Loss of seal integrity of the vials near storage temperatures of -80 °C would result in an ingress of cold dense gas into the vial headspace. After removal of the vials from storage at -80 °C, the rubber stoppers could regain their elastic properties and the vials would quickly reseal, thereby trapping the ingressed gas, which leads to overpressure in the vial headspace. Nondestructive laser-based headspace analysis was used to investigate the maintenance of container closure integrity as a function of the filling and capping/crimping process, storage and transport conditions, and vial/stopper designs. This analytical method is based on frequency modulation spectroscopy (FMS) and can be used for noninvasive headspace measurements of headspace pressure and headspace gas composition. Changes in the vial headspace composition and/or pressure are a clear marker for vials that have lost container closure integrity.

Lay Abstract: After storage of a live viral vaccine at -80 °C in glass vials closed with rubber stoppers, overpressure in some of the vials was observed, posing a serious safety problem for medical personnel as well as for the product. A working hypothesis to explain this phenomenon involved a possible container closure integrity issue at these low temperatures. The glass transition temperatures (T(g)) of many rubber stopper formulations are in the range -55 to -70 °C. At storage temperatures below T(g), the rubber stopper loses its elastic properties, resulting in compromised seal integrity of the vial and ingress of cold dense gas into the vial headspace. Upon thawing, the rubber stoppers regain their elastic properties and the vials quickly reseal, thereby trapping the ingressed gas, which leads to overpressure in the vial headspace. Nondestructive, laser-based headspace analysis, which is able to detect changes in headspace pressure and gas composition, was used to investigate the maintenance of container closure integrity. Changes in the vial headspace composition and/or pressure are a clear marker for vials that have lost container closure integrity.

Download full-text PDF

Source
http://dx.doi.org/10.5731/pdajpst.2012.00884DOI Listing

Publication Analysis

Top Keywords

container closure
24
vial headspace
24
closure integrity
20
-80 °c
16
rubber stoppers
16
temperatures rubber
16
rubber stopper
16
elastic properties
16
storage temperatures
12
seal integrity
12

Similar Publications

Biological semi-passive mine water treatment technologies are used in the mining industry as an alternative to or in conjunction with active treatment systems to remediate mine impacted water (MIW) containing nitrate and selenium oxyanions such as selenate and selenite. In semi-passive biological treatment systems, MIW is pumped through a saturated, porous media (either a gravel bed or waste rock) which provides ample surface area for biofilm growth and the creation of anoxic, subaqueous environments. Additional nutrients and carbon sources are pumped into the system to encourage the growth of microbes that biochemically reduce selenate and selenite to insoluble reduced Se species such as selenium nanoparticles (SeNP) by respiring selenate and selenite.

View Article and Find Full Text PDF

This study introduces a novel bilayer wound dressing that integrates a quaternized chitosan-polyacrylic acid (QCs-PAA) sponge as the top layer with electrospun nanofibers containing curcumin as the bottom layer. For the first time, QCs and PAA were combined in an 80:20 ratio through freeze-drying to form a porous sponge layer with ideal structural properties, including 83 ± 6 % porosity and pore diameters of 290 ± 12.5 μm.

View Article and Find Full Text PDF

Purpose Of The Review: This review examines the role of vascular access and closure management in cardiac electrophysiology (EP) procedures, emphasising their impact on patient outcomes and safety. It synthesises current evidence and highlights advancements, challenges, and opportunities in this critical area of EP practice.

Recent Findings: Ultrasound-guided vascular access has significantly reduced complications and improved success rates compared to traditional methods.

View Article and Find Full Text PDF

Practical advice in the development of a lyophilized protein drug product.

Antib Ther

January 2025

Formulation Development Group, Regeneron Pharmaceuticals Inc, 777 Old Saw Mill River Rd, Tarrytown, NY 10591, USA.

The development of lyophilized protein drug products is a critical and complex task in the pharmaceutical industry, requiring a comprehensive understanding of the myriad of factors affecting product quality, stability, and the efficiency and robustness of the lyophilization process. This review offers practical advice on the critical aspects of lyophilized protein drug product development. Practical considerations across both the early and late stages of development are discussed, underscoring the necessity of a strategic approach from initial development through to commercialization.

View Article and Find Full Text PDF

Electrospun robust, biodegradable, bioactive, and nanostructured sutures to accelerate the chronic wound healing.

Biofabrication

January 2025

College of Textiles & Clothing, Qingdao University, 308 Ningxia Road, Qingdao, Qingdao, Shandong, 266071, CHINA.

The design and development of advanced surgical sutures with appropriate structure and abundant bio-functions are urgently required for the chronic wound closure and treatment. In this study, an integrated technique routine combining modified electrospinning with hot stretching process was proposed and implemented to fabricate poly(L-lactic acid) (PLLA) nanofiber sutures, and the Salvia miltiorrhiza Bunge-Radix Puerariae herbal compound (SRHC) was encapsulated into PLLA nanofibers during the electrospinning process to enrich the biofunction of as-generated sutures. All the PLLA sutures loading without or with SRHC were found to exhibit bead-free and highly-aligned nanofiber structure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!