Objective: To assess the possibility of improving nocturnal glycemic control as well as meal glycemic response using closed-loop therapy in children aged <7 years.

Research Design And Methods: This was a randomized controlled crossover trial comparing closed-loop with standard open-loop insulin pump therapy performed in an inpatient clinical research center. Ten subjects aged <7 years with type 1 diabetes for >6 months treated with insulin pump therapy were studied. Closed-loop therapy and standard open-loop therapy were compared from 10:00 p.m. to 12:00 p.m. on 2 consecutive days. The primary outcome was plasma glucose time in range (110-200 mg/dL) during the night (10:00 p.m.-8:00 a.m.). Secondary outcomes included peak postprandial glucose levels, incidence of hypoglycemia, degree of hyperglycemia, and prelunch glucose levels.

Results: A trend toward a higher mean nocturnal time within target range was noted for closed- versus open-loop therapy, although not reaching statistical significance (5.3 vs. 3.2 h, P = 0.12). There was no difference in peak postprandial glucose or number of episodes of hypoglycemia. There was significant improvement in time spent >300 mg/dL overnight with closed-loop therapy (0.18 vs. 1.3 h, P = 0.035) and the total area under the curve of glucose >200 mg/dL (P = 0.049). Closed-loop therapy returned prelunch blood glucose closer to target (189 vs. 273 mg/dL on open loop, P = 0.009).

Conclusions: Closed-loop insulin delivery decreases the severity of overnight hyperglycemia without increasing the incidence of hypoglycemia. The therapy is better able to reestablish target glucose levels in advance of a subsequent meal. Younger children with type 1 diabetes may reap significant benefits from closed-loop therapy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3554296PMC
http://dx.doi.org/10.2337/dc12-1079DOI Listing

Publication Analysis

Top Keywords

closed-loop insulin
8
closed-loop therapy
8
closed-loop
4
therapy
4
insulin therapy
4
therapy improves
4
improves glycemic
4
glycemic control
4
control children
4
children aged
4

Similar Publications

Patients with type 1 diabetes and their physicians have long desired a fully closed-loop artificial pancreas (AP) system that can alleviate the burden of blood glucose regulation. Although deep reinforcement learning (DRL) methods theoretically enable adaptive insulin dosing control, they face numerous challenges, including safety and training efficiency, which have hindered their clinical application. This paper proposes a safe and efficient adaptive insulin delivery controller based on DRL.

View Article and Find Full Text PDF

This study evaluated a next-generation automated insulin delivery (AID) algorithm for Omnipod in type 1 and type 2 diabetes across multiple phases: 14-day run-in with usual therapy, 48-h AID use in a hotel setting (type 1 only), and up to 6 weeks of outpatient AID use. Participants did, or did not, deliver manual boluses at alternating periods. Twelve adults with type 1 diabetes completed the hotel phase; 9 of those 12 plus 8 adults with type 2 diabetes completed the subsequent outpatient phase.

View Article and Find Full Text PDF

To evaluate real-world outcomes in adults with type 1 diabetes initiating open-source automated insulin delivery systems (OS-AID). Adults with type 1 diabetes who commenced OS-AID, between May 2016 and April 2021, across 12 centers in the United Kingdom were included. Anonymized clinical data, collected during routine clinical care between December 2019 and November 2023, were submitted to a secure web-based tool within the National Health Service network.

View Article and Find Full Text PDF

Objectives: Type 1 diabetes (T1D) in pregnancy is challenging. This study explores how assisted hybrid closed-loop (HCL) therapy versus sensor-augmented pump therapy (SAPT) impacts quality of life (QoL) in pregnancy.

Methods: We interviewed 22 of 24 participants randomized to HCL therapy or SAPT in the Pregnancy Intervention with a Closed-Loop System study.

View Article and Find Full Text PDF

Technology usage and glycaemic outcomes in a single tertiary centre with an ethnically diverse and socioeconomically deprived cohort of children with type 1 diabetes mellitus.

Front Clin Diabetes Healthc

January 2025

Department of Endocrinology and Diabetes, Birmingham Women's and Children's NHS Foundation Trust, Birmingham, United Kingdom.

Background: The UK National Paediatric Diabetes Audit (NPDA) data reports disparities in Haemoglobin A1c (HbA1c) levels among children and young people (CYP) with Type 1 Diabetes (T1D), with higher levels in those of Black ethnic background and lower socioeconomic status who have less access to technology. We investigate HbA1c differences in a T1D cohort with higher than national average technology uptake where > 60% come from an ethnic minority and/or socioeconomically deprived population.

Design & Methods: Retrospective cross-sectional study investigating the influence of demographic factors, technology use, and socioeconomic status (SES) on glycaemic outcomes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!