Introduction: Transcatheter closure of patent ductus arteriosus (PDA) has replaced surgery in most institutions. Despite improvements in techniques and the devices available, closure of large PDA in very small infants remains a challenge.
Aim: To assess the challenges, feasibility, and efficacy of device closure of large PDA, in infants weighing ≤6 kg.
Materials And Methods: Analysis of device closure of a PDA was done in 61 infants ≤6 kg. Their ages, ranged from 9 days-12 months (mean 8.9 months), weight ranged from 2.2 to 6 kg (mean 5.3 kg), and PDA measured 3.2-8.7 mm (mean 4.8 mm). The fluoroscopy time was 3-18 minutes. The largest device used was 12 × 10 mm.
Results: Successful device placement was achieved in 60/61 infants (98.4%). Mild aortic obstruction occurred in 2 cases (3.3%), as the device got displaced towards the aorta after release. The device embolized in 2 cases (3.3%). In one it was retrieved by a novel method like fastening the screw in the aorta and was closed with a 4 × 6 ADO II. In the other infant, with a single kidney, died of uremia after device retrieval. Mild left pulmonary artery (LPA) obstruction occurred in one case (1.6%). Four cases (6.6%) had minor vascular complications. The postprocedure weight gain after 3 months was between 2.5 kg ± 250 mg.
Conclusions: Device closure of large PDA in infants weighing ≤6 kg with left ventricular failure is challenging but possible, safe and effective. Retrieval of embolized device could be tricky.