Purpose: To investigate the use of radiofrequency (RF) gating in conjunction with a paramagnetic contrast agent to reduce the specific absorption rate (SAR) and increase the blood-myocardium contrast in balanced steady-state free precession (bSSFP) 3D cardiac cine.

Materials And Methods: RF gating was implemented by synchronizing the RF-excitation with an external respiratory sensor (bellows), which could additionally be used for respiratory gating. For reference, respiratory-gated 3D cine images were acquired without RF gating. Free-breathing 3D cine images were acquired in eight healthy subjects before and after contrast injection (Gd-BOPTA) and compared to breath-hold 2D cine.

Results: RF-gated 3D cine reduced the SAR by nearly 40% without introducing significant artifacts while providing left ventricle (LV) measurements similar to those obtained with 2D cine. The contrast-to-noise ratio (CNR) was significantly higher for 3D cine compared to 2D cine, both before and after contrast injection; however, no statistically significant CNR increase was observed for the postcontrast 3D cine compared to the precontrast acquisitions.

Conclusion: Respiratory-triggered RF gating significantly reduces SAR in 3D cine acquisitions, which may enable a more widespread clinical use of 3D cine. Furthermore, CNR of 3D bSSFP cine is higher than of 2D and administration of Gd-BOPTA does not improve the CNR of 3D cine.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.23821DOI Listing

Publication Analysis

Top Keywords

cine
12
specific absorption
8
radiofrequency gating
8
cine images
8
images acquired
8
contrast injection
8
cine compared
8
gating
6
contrast-enhanced specific
4
absorption rate-efficient
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!