Background: The conventional right ventricle (RV) endocardial pacing leads QRS widening and desynchronization myocardial compromising ventricular function. With the need for stimulation less deleterious, RV septal pacing has been used more. Eventually have been reported higher thresholds and smaller R waves in the septal stimulation.

Objective: To compare the parameters of the septal and apical stimulation, intra-patient, if there are any differences that may affect the choice of the point of stimulation.

Methods: A prospective controlled study. We included 25 patients, 67.2±9 years, 10 (40%) women with indications for pacemaker for bradyarrhythmias. Etiologies were degenerative in nine (36%), Coronary disease in eight (32%), Chagas disease in seven (28%), and valve disease in one (4%) patient. Electrodes were active fixation and assessed the thresholds of command, impedance and R wave in uniand bipolar implant and after six months.

Results: The average acute threshold command, R wave and impedance unipolar / bipolar septais x apicais were respectively 0.73 x 0.73V and 0,74V x 0,78V; 10 x 9,9mV and 12,3 x 12,4mV; 579 x 621Ω and 611 x 629Ω. Comparisons between parameters with septal and apical two-tailed paired t-test showed a P > 0.1. After six months, the mean control thresholds, R wave impedances and unipolar/bipolar septais x apicais were respectively 0.5V x 0 72V and 0.71V x 0,87V; 11.4 x 9,5mV and 12x11,2mV; 423x426 Ω and 578x550 Ω, with P > 0.05, except compared to unipolar pacing threshold septal apical unipolar P 0.02.

Conclusion: Using intra-patient comparisons, no significant differences between electrophysiological parameters septal and apical pacing and there are no restrictions for choosing the right ventricular septal pacing.

Download full-text PDF

Source
http://dx.doi.org/10.5935/1678-9741.20120055DOI Listing

Publication Analysis

Top Keywords

septal apical
20
parameters septal
16
electrophysiological parameters
8
septal
8
septal pacing
8
septais apicais
8
apical
5
pacing
5
comparison electrophysiological
4
parameters
4

Similar Publications

Background Ventricular septal pacing has long been performed using a stylet during pacemaker implantation, but with the availability of guiding catheters, His bundle pacing and left bundle branch area pacing have also been performed. However, it is not known to what extent the tip load of the ventricular lead differs when a guiding catheter is used compared with a stylet alone. In this study, the tip load was measured for different stylet stiffness and guiding catheter geometries at sites where His bundle pacing and left bundle branch area pacing were assumed.

View Article and Find Full Text PDF

We report the case of a 3-year-old asymptomatic girl (12 kg, 96 cm) who was diagnosed with a large iatrogenic left ventricular pseudoaneurysm (LVP) on follow-up ultrasound, 14 months after apical muscular ventricular septal defect (VSD) closure with a 10 mm Amplatzer Muscular VSD occluder (Abbott, USA) due to device erosion. The LVP was successfully occluded using detachable Penumbra coils, with complete thrombo-exclusion confirmed at 12-month follow-up.

View Article and Find Full Text PDF

Background: In obstructive hypertrophic cardiomyopathy, myectomy improves symptoms, quality of life, and left ventricular (LV) outflow tract gradients. We prospectively evaluated the temporal changes in various echo parameters after myectomy.

Methods And Results: In 173 adults with obstructive hypertrophic cardiomyopathy (53±10 years, 63% men) who underwent myectomy between March 2017 and June 2020, clinical and blinded echo assessment (before and at 12±6 months follow-up) was performed prospectively (SPIRIT-HCM [Quality of Life and Functional Capacity Following Septal Myectomy in Obstructive Patients With Hypertrophic Cardiomyopathy]).

View Article and Find Full Text PDF

Complex ventricular tachycardias involving the fascicular system (fascicular ventricular tachycardias [FVTs]) can be challenging. In this review, we describe our approach to the diagnosis and ablation of these arrhythmias with 10 illustrative cases that involve (1) differentiation from supraventricular tachycardia; (2) assessment for atypical bundle branch reentry and other interfascicular FVTs; (3) examination of P1/P2 activation sequences in sinus rhythm, pacing, and tachycardia; and (4) entrainment techniques to establish the tachycardia mechanism and aid circuit localization. To summarize, 5 cases had prior ablation with 2 previously misdiagnosed as supraventricular tachycardia.

View Article and Find Full Text PDF

Unlabelled: This study investigates a previously unreported stress signal transduced as crosstalk between the cell wall integrity (CWI) pathway and the septation initiation network (SIN). Echinocandins, which target cell wall synthesis, are widely used to treat mycoses. Their efficacy, however, is species specific.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!