Purpose: To combine fluorine 19 ((19)F) magnetic resonance (MR) imaging and golden angle radial acquisition and to assess the feasibility of (19)F MR imaging golden angle-based tracking for catheter tracking applications and simultaneous three-dimensional (3D) intestinal tracking of ingested (19)F-labeled capsules in vivo.
Materials And Methods: Approval from the local ethical committee and informed consent from the subject were obtained. In vitro studies were performed to assess (19)F MR imaging golden angle-based tracking reliability with regard to temporal resolution and different tracking strategies (boundary condition-free tracking, composite image-based tracking, and model-based tracking). In vivo performance of the method was investigated in one healthy volunteer on 2 days. On study day 1, a duodenal catheter incorporating five (19)F-labeled capsules was administered nasally, and its 3D movement was tracked inside the stomach and esophagus. On study day 2, three (19)F-labeled capsules were swallowed, and intestinal movement was tracked.
Results: Simultaneous in vivo 3D tracking of multiple (19)F-labeled capsules was successfully performed without incorporation of boundary conditions at a temporal resolution of 252 msec. Incorporation of boundary conditions with composite image-based tracking and model-based tracking increased tracking reliability and enabled temporal resolution as high as 108 msec.
Conclusion: Use of (19)F MR imaging golden angle-based capsule tracking enables in vivo tracking of (19)F-labeled capsules and catheters at high temporal resolution. The presented method is applicable to physioanatomic studies of the gastrointestinal tract and shows potential for real-time tracking in interventional radiology.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1148/radiol.12120297 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!