Objective: The purpose of this study was to compare image quality of iterative reconstruction (IR) to filtered back projection (FBP) in low-dose computed tomography of the chest.

Methods: Forty-three consecutive patients were retrospectively enrolled. Eight series of images were reconstructed using FBP and 7 levels of IR in each subject. Image noise, signal-to-noise ratio (SNR), and SNR improvement were measured. Two radiologists evaluated subjective artifact, image artificiality, and subjective overall image quality with 4- or 5-point scales.

Results: Iterative reconstruction showed significantly lower image noise (135.5 ± 36.6 vs 219.9 ± 40.9) and higher SNR (0.36 ± 0.12 vs 0.21 ± 0.05) than FBP (P < 0.001). Signal-to-noise ratio improvement was 72.4% ± 44.9%. Subjective artifact of FBP was significantly higher than IR images (P < 0.001). Image artificiality of IR was significantly higher than that of FBP (P < 0.001). Overall, subjective image quality was poor in FBP and acceptable or good in IR.

Conclusions: With the use of IR, low-dose computed tomography of the chest would achieve less image noise and better image quality compared to the FBP.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0b013e318260ce39DOI Listing

Publication Analysis

Top Keywords

image quality
20
low-dose computed
12
computed tomography
12
iterative reconstruction
12
image noise
12
image
10
tomography chest
8
filtered projection
8
signal-to-noise ratio
8
subjective artifact
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!