Purpose: To compare central macular thickness (CMT) measurement on healthy patient using 3 different OCT devices by two operators.

Methods: Prospective, monocentricstudy. Right eye's central macular thickness (CMT) of 30 healthy patients has been measured three times using a time-domain (TD) OCT (Stratus OCT, Carl Zeiss Meditec, Dublin, Ca) and two spectral domain (SD) OCTs (Cirrus HD-OCT, Carl ZeissMeditec, Dublin, Ca) and 3D-OCT 1000 (Topcon, Tokyo, Japan) by two operators. Six measurements were taken randomly for each patient the same day.

Results: No significant difference between measurements obtained by the two operators has been observed, whatever the studied OCT. P value was 0.164, 0.193, and 0.147 for Stratus OCT, Cirrus HD-OCT and 3D-OCT, respectively. Mean CMT significantly differed from instrument to instrument (P < 0.001) and was, respectively, 197 μm, 254 μm, and 236 μm using Stratus OCT, Cirrus HD-OCT, and 3D-OCT 1000. Using Cirrus OCT and 3D-OCT 1000, CMT was, respectively, 57 μm and 39 μm thicker than using Stratus OCT (P < 0.05).

Conclusions: Whatever the OCT device, on healthy patients CMT was not operator dependent. CMT measurements obtained by SD-OCTs are greater than those obtained by TD-OCT. These data imply that the different OCT devices cannot be used interchangeably in clinical monitoring.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3432357PMC
http://dx.doi.org/10.1100/2012/842795DOI Listing

Publication Analysis

Top Keywords

stratus oct
16
central macular
12
macular thickness
12
cirrus hd-oct
12
3d-oct 1000
12
oct
11
measured three
8
thickness cmt
8
oct devices
8
healthy patients
8

Similar Publications

Concurrent electrocardiogram (ECG) and electroencephalogram (EEG) recording both ictally and interictally has significant value in the comprehensive management of epilepsy. This review highlights the diagnostic utility of simultaneous ECG and EEG monitoring in differentiating between epileptic and cardiac events, detecting cardiac abnormalities, and identifying autonomic dysfunction. The critical role of this combined approach to defining the mechanisms underlying cardiac morbidity and sudden cardiac death in patients with epilepsy and in guiding therapeutic interventions is underscored.

View Article and Find Full Text PDF

Background: Previous study of spinal neurotomy procedures indicates that stringent block selection improves outcomes. However, this pattern is not established for sacral lateral branch radiofrequency neurotomy (SLBRFN). Few SLBRFN studies have used stringent block selection criteria such as ≥80 % pain reduction following dual sacral lateral branch blocks (SLBB).

View Article and Find Full Text PDF

Purpose: To concurrently assess and compare the fixation efficacy, invasiveness, and fusion potential of a posterior integrated transfixation cage system to the posterolateral threaded implant and lateral triangular rod systems, in a cadaveric model.

Methods And Materials: Twelve (12) cadaveric sacroiliac joint specimens were utilized and tested within the single-leg stance multidirectional pure moment bending model. Each specimen was tested in the intact, destabilized, treated (using posterior, posterolateral, and lateral systems), and post-fatigue conditions by applying 0 to ± 7.

View Article and Find Full Text PDF
Article Synopsis
  • This review looks at how augmented reality (AR) and virtual reality (VR) can help people who have trouble hearing communicate better.
  • Researchers found 22 studies focused mainly on nonverbal ways to communicate, like using sign language and visual signals.
  • The studies suggest that AR and VR can help teach sign language and improve thinking and speech skills for those who are hard of hearing, but more research is needed to explore their full potential.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!