Purpose: To determine the effects of imager source and panel positioning uncertainties on the accuracy of dual intensity-based 2D∕3D image registration of cranial images.

Methods: An open source 2D∕3D image registration algorithm has been developed for registration of two orthogonal x-rays to a 3D volumetric image. The initialization files of the algorithm allow for nine degrees of freedom system calibration including x, y, z positions of the source and panel, and three rotational degrees of freedom of the panel about each of the three translational axes. A baseline system calibration was established and a baseline 2D∕3D registration between two orthogonal x-rays and the volumetric image was determined. The calibration file was manipulated to insert errors into each of the nine calibration variables of both imager geometries. Rigid six degrees of freedom registrations were iterated for each panel or source positional error over a range of predetermined calibration errors to determine the resulting error in the registration versus the baseline registration due to the manipulated error of the panel or source calibration.

Results: Panel and source translational errors orthogonal to the imager∕panel axis introduced the greatest errors in the registration accuracy (4.0 mm geometric error results in up to 2.7 mm registration error). Panel rotation about the imaging direction also resulted in errors of the registration (2.0° geometric error results in up to 1.7° registration error). Differences in magnification and panel tilt and roll, i.e., source and∕or panel translation along the imaging direction and panel rotations about the orthogonal axes had minimal effects on the registration accuracy (below 0.3 mm and 0.2° registration error).

Conclusions: While five of the nine imaging system variables were found to have a considerable effect on 2D∕3D registration accuracy of cranial images, the other four variables showed minimal effects. Vendors typically provide simplified calibration procedures which aim to remove encountered geometric uncertainties by accounting for two panel translations. This study shows that at least the five relevant positional variables should be separately calibrated, if accurate alignment is required for 2D∕3D registration.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4742866DOI Listing

Publication Analysis

Top Keywords

registration
15
panel
12
source panel
12
2d∕3d image
12
image registration
12
degrees freedom
12
2d∕3d registration
12
panel source
12
registration accuracy
12
source
8

Similar Publications

Background:  Ultrasonographic measurement of optic nerve sheath diameter (ONSD) has been proposed as a non-invasive, bedside method to detect raised intracranial pressure (ICP) in various clinical settings. We aimed to correlate the ONSD obtained by ultrasonography (USG) with the ONSD obtained by magnetic resonance imaging (MRI) and to find its measurement accuracy.

Methodology:  A prospective double-blind study was carried out by performing ocular ultrasounds on 32 patients with clinical features of intracranial hypertension.

View Article and Find Full Text PDF

Background: Kentucky is one of seven states with high, sustained rural HIV transmission tied to injection drug use. Expanding access to pre-exposure prophylaxis (PrEP) has been endorsed as a key HIV prevention strategy; however, uptake among people who inject drugs (PWID) has been negligible in rural areas. Syringe services programs (SSPs) have been implemented throughout Kentucky's Appalachian region, providing an important opportunity to integrate PrEP services.

View Article and Find Full Text PDF

Aim: To identify existing public knowledge regarding diabetes and diabetes-related services offered to persons living with diabetes in the City of Malmö.

Methods: A literature review of City of Malmö's website, public statistics, School health documentation, job databases, education programs, local newspaper, Swedish National Diabetes Register, and was performed in 2020.

Results: We identified political decisions about diabetes nurses in home care, financing a project about diabetes complications, and funding support in schools for designated children.

View Article and Find Full Text PDF

Introduction: The incidence of pediatric tracheostomy is on the rise. More children are undergoing tracheostomy at a younger age and living longer and cared for at home. Caring for children with tracheostomy affects the caregivers' Quality of Life (QOL) and caregiver burden.

View Article and Find Full Text PDF

Objective: To review the efficacy and safety of low-dose versus standard-dose enzalutamide, apalutamide or darolutamide treatment for metastatic prostate cancer.

Methods And Analysis: Keyword searches in MEDLINE and EMBASE up to 1 June 2023, with forward and backward citation searches of potentially relevant studies. Studies were included if primary outcome data were reported for patients with metastatic prostate cancer who had received reduced doses of enzalutamide, apalutamide or darolutamide.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!