An aqueous normal phase (ANP) liquid chromatography coupled with a hybrid quadrupole time-of-flight mass spectrometry (ANP-LC-micrOTOFQ) method was used for the determination of zanamivir in human serum. Zanamivir was extracted with methanol from protein-precipitated human serum samples and further purified with SCX solid-phase extraction cartridges. Scherzo SM-C18, Agilent Zorbax SB-Aq, Cogent Diamond Hydride, Cogent Bidentate and Luna HILIC columns were compared and optimized for the retention and separation of zanamivir and the Luna HILIC and Diamond Hydride columns exhibited the best retention of zanamivir. The former provided a shorter retention time, a sharper peak and relatively high sensitivity, whereas the latter exhibited a longer retention time and less matrix interference. The analytical range of the calibration curve was between 5 and 1000 ng/mL.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3465458 | PMC |
http://dx.doi.org/10.1016/j.jchromb.2012.08.020 | DOI Listing |