Assessment of on-screen measurements, magnification, and calibration in digital radiography.

J Am Vet Med Assoc

Aspen Meadow Veterinary Specialists, Longmont, CO 80501, USA.

Published: September 2012

Objective: To compare calibration methods for digital radiography in terms of measurement accuracy and interobserver variability.

Design: Prospective study.

Sample: Digital radiographic images of a 155-mm-long Steinmann pin.

Procedures: Measurement of pin length on digital radiographs was determined with a 25.4-mm-diameter calibration ball and commercially available software program via 3 calibration methods (ie, no calibration, autocalibration, and manual calibration). Digital radiographs of the calibration ball and pin were obtained with each placed at various vertical heights from the table (7 heights) and horizontal distances from the center of the beam (4 distances). Measurements of pin length on digital radiographs were made by 4 observers who were blinded to the orientation of the calibration ball and pin.

Results: Pin lengths obtained by each calibration method were significantly different from each other and from the true value. Manual calibration was the most accurate. There was no significant interobserver variability in measurements. There was no significant change in measurements when the calibration ball was moved horizontally, but pin length measurements changed significantly when the ball was moved vertically (away from the table) with an approximate magnification error of 1% per centimeter of distance between the calibration ball and pin.

Conclusions And Clinical Relevance: For digital radiography, manual calibration is recommended to achieve the most accurate measurements. Ideally, the calibration ball should be placed at the same vertical height as the object to be measured; however, if this cannot be achieved, the magnification error can be expected to be approximately 1% per centimeter of distance.

Download full-text PDF

Source
http://dx.doi.org/10.2460/javma.241.6.782DOI Listing

Publication Analysis

Top Keywords

calibration ball
24
calibration
14
digital radiography
12
pin length
12
digital radiographs
12
manual calibration
12
calibration digital
8
calibration methods
8
length digital
8
ball moved
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!