Micro particle image velocimetry (µPIV) is a common method to assess flow behavior in blood microvessels in vitro as well as in vivo. The use of red blood cells (RBCs) as tracer particles, as generally considered in vivo, creates a large depth of correlation (DOC), even as large as the vessel itself, which decreases the accuracy of the method. The limitations of µPIV for blood flow measurements based on RBC tracking still have to be evaluated. In this study, in vitro and in silico models were used to understand the effect of the DOC on blood flow measurements using µPIV RBC tracer particles. We therefore employed a µPIV technique to assess blood flow in a 15 µm radius glass tube with a high-speed CMOS camera. The tube was perfused with a sample of 40% hematocrit blood. The flow measured by a cross-correlating speckle tracking technique was compared to the flow rate of the pump. In addition, a three-dimensional mechanical RBC-flow model was used to simulate optical moving speckle at 20% and 40% hematocrits, in 15 and 20 µm radius circular tubes, at different focus planes, flow rates and for various velocity profile shapes. The velocity profiles extracted from the simulated pictures were compared with good agreement with the corresponding velocity profiles implemented in the mechanical model. The flow rates from both the in vitro flow phantom and the mathematical model were accurately measured with less than 10% errors. Simulation results demonstrated that the hematocrit (paired t tests, p = 0.5) and the tube radius (p = 0.1) do not influence the precision of the measured flow rate, whereas the shape of the velocity profile (p < 0.001) and the location of the focus plane (p < 0.001) do, as indicated by measured errors ranging from 3% to 97%. In conclusion, the use of RBCs as tracer particles makes a large DOC and affects the image processing required to estimate the flow velocities. We found that the current µPIV method is acceptable to estimate the flow rate on the condition that the measurement takes place at the equatorial plane of the vessel. Otherwise, it is not an appropriate method to estimate the shape of the velocity profile.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0967-3334/33/10/1585DOI Listing

Publication Analysis

Top Keywords

blood flow
16
flow
12
tracer particles
12
flow rate
12
velocity profile
12
rbcs tracer
8
flow measurements
8
flow rates
8
velocity profiles
8
shape velocity
8

Similar Publications

Langerhans cell histiocytosis in children: the value of ultrasound in diagnosis and follow-up.

BMC Med Imaging

January 2025

Department of Ultrasound Medicine, First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China.

Background: Langerhans cell histiocytosis (LCH) is a rare disease, most prevalent in children. Ultrasound is a noninvasive, cheap, and widely available technique. However, systematic elucidation of sonographic features of LCH and treatment related follow-up are relatively few, resulting in overall underestimation of the clinical value of ultrasound in diagnosing and monitoring LCH.

View Article and Find Full Text PDF

Carcinoembryonic antigen (CEA) and C-reactive protein (CRP) are biomacromolecules known as cancer and inflammatory markers. Thus, they play a crucial role in early cancer diagnosis, post-treatment recurrence detection, and tumor risk assessment. This paper describes the development of an ultrasensitive and selective imprinted paper-based analytical device (PAD) as impedance sensor for determination of CEA and CRP in serum samples for point-of-care testing (POCT).

View Article and Find Full Text PDF

Reduced cerebral blood flow occurs early in the development of Alzheimer's disease (AD), but the factors producing this reduction are unknown. Here, we ask whether genetic and lifestyle risk factors for AD-the ε4 allele of the Apolipoprotein (APOE) gene, and physical activity-can together produce this reduction in cerebral blood flow which leads eventually to AD. Using in vivo two-photon microscopy and haemodynamic measures, we record neurovascular function from the visual cortex of physically active or sedentary mice expressing APOE3 and APOE4 in place of murine APOE.

View Article and Find Full Text PDF

Aptamer-Antibody Birecognized Sandwich SERRS Biosensor in Accurate and Rapid Identification of Intraoperative Parathyroid Hormone.

Anal Chem

January 2025

State Key Laboratory of Physical Chemistry of Solid Surfaces, College of Chemistry and Chemical Engineering, College of Energy, Discipline of Intelligent Instrument and Equipment, Cancer Center and Department of Breast and Thyroid Surgery, Department of Ultrasound, Xiang'an Hospital of Xiamen University, School of Medicine, Laboratory Animal Center Xiamen University, Xiamen University, Xiamen 361005, China.

With the increasing incidence of thyroid cancer worldwide and the increasing demand for surgery, the risk of parathyroid injury is also increasing, which will lead to postoperative hypoparathyroidism (HP) and hypocalcemia. In order to improve the quality of life of patients after surgery, there is an urgent need to develop a novel platform that can identify the parathyroid gland immediately during surgery. The parathyroid gland promotes the increase of blood calcium concentration by secreting parathyroid hormone (PTH).

View Article and Find Full Text PDF

Background: Comprehensive management of acute coronary syndrome (ACS) requires seamless treatment across institutions, including intensive care centers and local clinics. However, maintaining guideline-directed medical therapy remains challenging. One promising option to improve the situation may be the implementation of regional collaborative clinical pathways.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!