Device life of the Provox Vega voice prosthesis.

Eur Arch Otorhinolaryngol

Speech Pathology Department, Princess Alexandra Hospital, Brisbane, QLD, 4102, Australia.

Published: March 2013

Device life of the Provox Vega Indwelling voice prosthesis is as yet untested outside Europe. The current study examined device life and reasons for replacement within an Australian clinical setting. Twenty-three participants were monitored for device life and reasons for replacement. Main outcome measure was days to failure of initial device. Average device life and reasons for replacement were secondary measures. Initial device life data revealed 67 % had functioning devices at 3 months, 52 % at 6 months and 29 % at 12 months. Average device life was 207 days (median of 222). The majority of devices (97 %) failed due to leakage through the prosthesis. The Provox Vega Indwelling voice prosthesis had favourable device life in this cohort of patients and in comparison to European data. Reasons for replacement were consistent with international literature.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3608863PMC
http://dx.doi.org/10.1007/s00405-012-2154-9DOI Listing