Aim: To assess the visual outcomes and possible risk factors associated with axis alignment and rotational stability after implantation of Toric implantable collamer lens (TICL) for the correction of high myopic astigmatism.
Methods: In this prospective, nonrandomized clinical study, 54 consecutive eyes of 29 patients with high myopic astigmatism received TICL implantation. To evaluate postoperative axis deviation from the intended axis, a digital anterior segment photograph was taken. The ultrasound biomicroscopy(UBM) was used to observe footplate-position.
Results: After mean follow-up of 8.6 months, mean manifest refractive cylinder (MRC) decreased 79.3% from (-1.88±1.49)D preoperatively to (0.39±0.61)D postoperatively. MRC within 1.00 D occurred in 68.5% (37/54) of eyes, whereas 48.1% (26/54) had MRC within 0.50 D. Mean manifest refraction spherical equivalent (MRSE) changed from (-12.08±4.22)D preoperatively to (-0.41±0.61)D postoperatively. Uncorrected binocular vision of 20/20 or better occurred in 72.2% (39/54) of patients compared with binocular best-corrected visual acuity (BCVA) of 20/20 or better in 44.4% (24/54) preoperatively. The mean difference between intended and achieved TICL axes was (6.96±8.37)°. Footplates of TICLs were in the ciliary sulcus in 22 eyes (46.3%), below the ciliary sulcus in 32 eyes (53.7%). The angle of TICL rotation had significant correlation with the footplates-position (t=2.127; P=0.045) and the postoperative TICL vaulting (r=-0.516; P=0.000).
Conclusion: The results of our study further support the safety, efficacy and predictability of TICL for the correct high myopic astigmatism. The footplate-position of TICL and vault value should be taken into consideration as two possible risks factors for TICL rotation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3428541 | PMC |
http://dx.doi.org/10.3980/j.issn.2222-3959.2012.04.10 | DOI Listing |
Am J Ophthalmol
December 2024
State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Guangzhou, China. Electronic address:
Purpose: To investigate the ability to quantify fundus curvature and detect posterior staphyloma using widefield optical coherence tomography (OCT).
Design: Cross-sectional diagnostic evaluation.
Subjects And Participants: 205 highly myopic eyes of 205 participants.
Am J Ophthalmol
December 2024
Clinical College of Ophthalmology, Tianjin Medical University, Tianjin, China; School of Medicine, Nankai University, Tianjin, China; Tianjin Eye Hospital, Tianjin Key Laboratory of Ophthalmology and Visual Science, Tianjin Eye Institute, Nankai University Affiliated Ophthalmology Hospital, Tianjin, China; Nankai Eye Institute, Nankai University, Tianjin, China. Electronic address:
Purpose: To analyze the effect of individual parameters on the postoperative refractive outcomes of small incision lenticule extraction in myopic eyes using machine learning methods.
Design: Retrospective Clinical Cohort Study METHODS: We included 477 patients (922 eyes) of small incision lenticule extraction at Tianjin Ophthalmology Hospital and divided the patients into two groups to analyzed the factors affecting postoperative refractive outcomes based on the label of postoperative spherical equivalent (SE) ≤ -0.50D.
Retina
October 2024
Division of Ophthalmology, Department of Visual Sciences, Nihon University School of Medicine, Nihon, Japan.
Purpose: To investigate the impact of foveal glial tissue on the anatomical and functional results after macular hole (MH) surgery.
Methods: This study included 141 consecutive eyes that underwent successful vitrectomy for full-thickness MH between January 2015 and December 2022. The best-corrected visual acuity (BCVA) and the length of outer retinal defects were evaluated preoperatively and at 6 months postoperatively.
Vestn Oftalmol
December 2024
Krasnov Research Institute of Eye Diseases, Moscow, Russia.
Purpose: This study compares the functional outcomes of correction using two different types of aberrational extended depth-of-focus (EDOF) intraocular lenses (IOLs).
Material And Methods: The study was conducted in two standardized groups (20 patients, 20 eyes in each group). Patients in group 1 were implanted EDOF IOL I, in group 2 - EDOF IOL II.
Sci Rep
December 2024
Department of Ophthalmology, Jinshan Hospital of Fudan University, 1508 Longhang Road, Jinshan District, Shanghai, China.
To observe the structural changes of retina and choroid in patients with different degrees of myopia. We recruited 219 subjects with different degrees of myopia for best corrected visual acuity, computer refraction, intraocular pressure, axial length (AL), optical coherence tomography (OCT) imaging, and other examinations. Central macular retinal thickness (CRT), subfoveal choroidal thickness (SFCT), nasal retinal thickness (NRT), temporal retinal thickness (TRT), nasal choroidal thickness (NCT) and temporal choroidal thickness (TCT) were measured by optical coherence tomography.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!