A method to localize RF B₁ field in high-field magnetic resonance imaging systems.

IEEE Trans Biomed Eng

Department of Electrical and Computer Engineering and the Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN 55455, USA.

Published: December 2012

In high-field magnetic resonance imaging (MRI) systems, B₀ fields of 7 and 9.4 T, the RF field shows greater inhomogeneity compared to clinical MRI systems with B₀ fields of 1.5 and 3.0 T. In multichannel RF coils, the magnitude and phase of the input to each coil element can be controlled independently to reduce the nonuniformity of the RF field. The convex optimization technique has been used to obtain the optimum excitation parameters with iterative solutions for homogeneity in a selected region of interest. The pseudoinverse method has also been used to find a solution. The simulation results for 9.4- and 7-T MRI systems are discussed in detail for the head model. Variation of the simulation results in a 9.4-T system with the number of RF coil elements for different positions of the regions of interest in a spherical phantom are also discussed. Experimental results were obtained in a phantom in the 9.4-T system and are compared to the simulation results and the specific absorption rate has been evaluated.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4041104PMC
http://dx.doi.org/10.1109/TBME.2012.2208965DOI Listing

Publication Analysis

Top Keywords

mri systems
12
high-field magnetic
8
magnetic resonance
8
resonance imaging
8
systems b₀
8
b₀ fields
8
94-t system
8
method localize
4
localize b₁
4
b₁ field
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!