Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The efficacy of tuned defibrillation waveforms versus the nominal fixed-tilt waveform has been previously studied. However, the optimal membrane time constant for tuning was not known. The POWER (Pulsewidth Optimized Waveform Evaluation tRial) trial was designed to determine the optimal membrane time constant for programming "tuned" biphasic waveforms.
Methods: This acute, multicenter study included 121 implantable cardioverter-defibrillator/cardiac resynchronization therapy defibrillator patients who were randomized at implant to any two of the three membrane time constant waveforms (2.5, 3.5, and 4.5 ms). Fixed pulse widths were programmed using the measured high voltage shock impedance. The defibrillation threshold (DFT) estimates were obtained using a hybrid protocol starting with an upper limit of vulnerability estimate followed by a step-up/step-down ventricular fibrillation induction process.
Results: DFT voltage was significantly lower using 3.5- and 4.5-ms waveforms as compared to the 2.5-ms waveform (P = 0.004 and 0.035, respectively). DFT voltage with both 3.5- and 4.5-ms waveforms was ≤ that obtained with the 2.5-ms waveform in 78.5% of the cases. The mean difference in DFT voltage using the 3.5-ms waveform and the 4.5-ms waveform was not significant (P = 0.4). However, the 3.5-ms waveform gave a lower DFT than the 4.5-ms waveform in 19 patients although the reverse was true in only nine (P = 0.02 not significant for multiple comparisons).
Conclusions: The use of a 3.5- or 4.5-ms time constant-based waveforms had lower DFTs when compared to the 2.5-ms waveform. This study suggests that the first defibrillation attempt at implantation should be with 3.5- or 4.5-ms time constant-based waveforms. The 3.5-ms-based waveform trended toward the best choice.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8159.2012.03500.x | DOI Listing |