Aim of the study was to assess the accuracy and precision of a BIS device and the relative contribution of BIS beyond the anthropometric parameters. The output of the Impedimed device (SFB7) and the relative contribution of height, weight, age, sex and resistance values at zero and infinite frequency (Rzero and Rinf respectively) to the prediction of total body water (TBWd, deuterium space), of extracellular fluid (ECFbr, sodium bromide space) and of fat mass (FMDXA) were assessed in 116 subjects (32 healthy subjects and 84 patients with disorders of body composition). Using a repeated randomization procedure, new equations for TBW, ECF and FM were derived. The SFB7 gave measures of determination similar to those obtained with equations that included only anthropometric data. The SFB7, but not the newly derived regression equations, underestimated TBW and ECF by 3.82±3.37 (mean±SD) and by 0.93±2.62l and overestimated FM by 6.55±3.86kg. Nine of 16 patients with ECF overload as detected by ECFbr were also detected by BIS. BIS measurements contribute marginally but not significantly beyond anthropometric data to the prediction of TBW, ECF and FM, either in healthy subjects or in patients with disturbed body composition.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.medengphy.2012.07.006DOI Listing

Publication Analysis

Top Keywords

tbw ecf
12
relative contribution
8
healthy subjects
8
subjects patients
8
body composition
8
anthropometric data
8
body
5
bis
5
critical analysis
4
analysis body
4

Similar Publications

Objective: To assess whether bioimpedance spectroscopy analysis (BIA) can be used as a tool to guide postoperative fluid management in patients undergoing cardiac surgery.

Design: An observational study.

Setting: A single tertiary hospital.

View Article and Find Full Text PDF

Objective: To compare bioelectrical impedance analysis (BIA)-derived parameters in healthy volunteers and critically ill patients and to assess its prognostic value in an ICU patient cohort.

Design: Retrospective, observational data analysis.

Setting: Single centre, tertiary-level ICU (Ziekenhuis Netwerk Antwerpen, ZNA Stuivenberg Hospital).

View Article and Find Full Text PDF

Background: Spleen deficiency-water dampness symptom is closely related to body fluid-mediated organism metabolism and circulation. However, previous clinical evaluation of spleen deficiency-water dampness model was based only on body weight, D-xylose excretion rate, serum gastrin content, etc. Therefore, we established a large sample of normal rats and model rats experiment to verify the scientific nature of bio-impedance measuring body fluid indexes for evaluation of the modeling state.

View Article and Find Full Text PDF

Purpose: The purpose of this cross-sectional evaluation was to determine the associations between raw bioelectrical impedance and cardiometabolic health parameters in a sample of young non-Hispanic White and African American adults.

Methods: A total of 96 (female: 52, male: 44) non-Hispanic White ( n = 45) and African American adults ( n = 51) between the ages of 19 and 37 yr (22.7 ± 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!