Purpose: Most methods to estimate patient dose from computed tomography (CT) exams have been developed based on fixed tube current scans. However, in current clinical practice, many CT exams are performed using tube current modulation (TCM). Detailed information about the TCM function is difficult to obtain and therefore not easily integrated into patient dose estimate methods. The purpose of this study was to investigate the accuracy of organ dose estimates obtained using methods that approximate the TCM function using more readily available data compared to estimates obtained using the detailed description of the TCM function.

Methods: Twenty adult female models generated from actual patient thoracic CT exams and 20 pediatric female models generated from whole body PET∕CT exams were obtained with IRB (Institutional Review Board) approval. Detailed TCM function for each patient was obtained from projection data. Monte Carlo based models of each scanner and patient model were developed that incorporated the detailed TCM function for each patient model. Lungs and glandular breast tissue were identified in each patient model so that organ doses could be estimated from simulations. Three sets of simulations were performed: one using the original detailed TCM function (x, y, and z modulations), one using an approximation to the TCM function (only the z-axis or longitudinal modulation extracted from the image data), and the third was a fixed tube current simulation using a single tube current value which was equal to the average tube current over the entire exam. Differences from the reference (detailed TCM) method were calculated based on organ dose estimates. Pearson's correlation coefficients were calculated between methods after testing for normality. Equivalence test was performed to compare the equivalence limit between each method (longitudinal approximated TCM and fixed tube current method) and the detailed TCM method. Minimum equivalence limit was reported for each organ.

Results: Doses estimated using the longitudinal approximated TCM resulted in small differences from doses obtained using the detailed TCM function. The calculated root-mean-square errors (RMSE) for adult female chest simulations were 9% and 3% for breasts and lungs, respectively; for pediatric female chest and whole body simulations RMSE were 9% and 7% for breasts and 3% and 1% for lungs, respectively. Pearson's correlation coefficients were consistently high for the longitudinal approximated TCM method, ranging from 0.947 to 0.999, compared to the fixed tube current value ranging from 0.8099 to 0.9916. In addition, an equivalence test illustrated that across all models the longitudinal approximated TCM is equivalent to the detailed TCM function within up to 3% for lungs and breasts.

Conclusions: While the best estimate of organ dose requires the detailed description of the TCM function for each patient, extracting these values can be difficult. The presented results show that an approximation using available data extracted from the DICOM header provides organ dose estimates with RMSE of less than 10%. On the other hand, the use of the overall average tube current as a single tube current value was shown to result in poor and inconsistent estimates of organ doses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3422360PMC
http://dx.doi.org/10.1118/1.4736807DOI Listing

Publication Analysis

Top Keywords

tube current
40
tcm function
36
detailed tcm
32
tcm
17
fixed tube
16
organ dose
16
longitudinal approximated
16
approximated tcm
16
organ doses
12
dose estimates
12

Similar Publications

Rapid Preparation of Collagen/Red Blood Cell Membrane Tubes for Stenosis-Free Vascular Regeneration.

ACS Nano

January 2025

Shenzhen Key Laboratory of Smart Healthcare Engineering, Guangdong Provincial Key Laboratory of Advanced Biomaterials, Department of Biomedical Engineering, Southern University of Science and Technology, No. 1088 Xueyuan Road, Nanshan District, Shenzhen, Guangdong 518055, PR China.

Extracellular matrix (ECM)-based small-diameter vascular grafts (SDVGs, inner diameter (ID) < 6 mm) hold great promise for clinical applications. However, existing ECM-based SDVGs suffer from limited donor availability, complex purification, high cost, and insufficient mechanical properties. SDVGs with ECM-like structure and function, and good mechanical properties were rapidly prepared by optimizing common materials and preparation, which can improve their clinical prospects.

View Article and Find Full Text PDF

Background: Airway management is a critical component of the care of patients experiencing cardiac arrest, but data from randomized trials on the use of video vs direct laryngoscopy for intubation in the setting of cardiac arrest are limited. Current AHA guidelines recommend placement of an endotracheal tube either during CPR or shortly after return of spontaneous circulation but do not provide guidance around intubation methods, including the choice of laryngoscope.

Research Question: Does use of video laryngoscopy improve the incidence of successful intubation on the first attempt, compared to use of direct laryngoscopy, among adults undergoing tracheal intubation after experiencing cardiac arrest?

Study Design And Methods: This secondary analysis of the Direct versus Video Laryngoscope (DEVICE) trial compared video laryngoscopy versus direct laryngoscopy in the subgroup of patients who were intubated following cardiac arrest.

View Article and Find Full Text PDF

Enhanced Recovery After Surgery for patients undergoing radical cystectomy: Surgeons' perspectives and recommendations ten years after its implementation.

Eur J Surg Oncol

December 2024

Oncologic Minimally Invasive Urology and Andrology Unit, Department of Experimental and Clinical Medicine, Careggi Hospital, University of Florence, Florence, Italy.

Background And Objectives: Enhanced Recovery After Surgery (ERAS) guidelines for Radical Cystectomy (RC) were published over ten years ago. Aim of this systematic review is to update ERAS recommendations for patients undergoing RC and to give an expert opinion on the relevance of each single ERAS item.

Methods: A systematic review was performed to identify the impact of each single ERAS item on RC outcomes.

View Article and Find Full Text PDF

Background: Multianalyte machine learning (ML) models can potentially identify previously undetectable wrong blood in tube (WBIT) errors, improving upon current single-analyte delta check methodology. However, WBIT detection model performance has not been assessed in a real-world, low-prevalence context. To estimate real-world positive predictive values, we propose a methodology to assess WBIT detection models by evaluating the impact of missing data and by using a "low prevalence" validation data set.

View Article and Find Full Text PDF

Dynamic Digital Radiography (DDR) in the Diagnosis of a Diaphragm Dysfunction.

Diagnostics (Basel)

December 2024

Postgraduation School in Radiodiagnostics, Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy.

Dynamic digital radiography (DDR) is a recent imaging technique that allows for real-time visualization of thoracic and pulmonary movement in synchronization with the breathing cycle, providing useful clinical information. A 46-year-old male, a former smoker, was evaluated for unexplained dyspnea and reduced exercise tolerance. His medical history included a SARS-CoV-2 infection in 2021.

View Article and Find Full Text PDF