Background: This study was performed to analyse the reasons for explantation/exchange of intraocular lenses (IOL), which had originally been implanted for the correction of aphakia during cataract extraction.

Material And Methods: All cases with IOL explantation, which had been performed at one institution between 1/2008 and 12/2009 were analysed retrospectively.

Results: A total of 105 eyes of 100 patients were analysed. The median time interval between implantation and explantation of the IOL was 5.9 years (min. 0, max. 29.6). The most frequent cause for the intervention was subluxation/dislocation of the implant in 55.2% of cases. This group comprised 21% of cases with subluxation within the capsular bag in pseudoexfoliation syndrome. Other reasons were optical problems/incorrect IOL power (21%), calcification of hydrophilic acrylic IOL (7.6%), corneal decompensation associated with an anterior chamber lens (4.8%), and single cases with varying problems.

Conclusion: The reasons for IOL exchange presented in this study are comparable to those of other series in the literature. Explantations due to optical problems may gain weight in the future due to a rise in refractive procedures and demands.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-0031-1299520DOI Listing

Publication Analysis

Top Keywords

iol
6
[reasons exchange
4
exchange explantation
4
explantation intraocular
4
intraocular lenses]
4
lenses] background
4
background study
4
study performed
4
performed analyse
4
analyse reasons
4

Similar Publications

Introduction: In response to the increasing rates of induction of labor (IOL), outpatient IOL has emerged as a potential approach to enhance women's satisfaction while reducing costs and staffing requirements. There is a growing interest in oral misoprostol as an outpatient IOL method, particularly in the Nordic region. This study aims to evaluate the clinical outcomes and feasibility of implementing IOL with oral misoprostol as an outpatient procedure.

View Article and Find Full Text PDF

Editorial: Insights in glaucoma: 2023.

Front Ophthalmol (Lausanne)

January 2025

Singapore National Eye Centre, Singapore Eye Research Institute, Singapore, Eye-ACP, Duke-NUS Medical School, Singapore, Singapore.

View Article and Find Full Text PDF

Introduction: Iron overload (IOL) accumulates in myelodysplastic syndromes (MDS) from expanded erythropoiesis and transfusions. Somatic mutations (SM) are frequent in MDS and stratify patient risk. MDS treatments reversing or limiting transfusion dependence are limited.

View Article and Find Full Text PDF

To analyze the refractive accuracy of a novel swept-source optical coherence biometer (SS-OCT), that uses individual refractive indices to measure axial length, in short and long eyes implanted with monofocal intraocular lenses (IOLs). This retrospective comparative study considered eyes with short axial length (AL) (< 22.5 mm) or long AL (> 26 mm) bilaterally implanted with the Acrysof IQ monofocal IOL.

View Article and Find Full Text PDF

Purpose: To evaluate visual outcomes and quality of vision following bilateral implantation of a hydrophobic acrylic intraocular lens (IOL) in eyes targeted for emmetropia.

Methods: This was a prospective, single arm study. Subjects were bilaterally implanted with the Clareon PanOptix IOL and evaluated at 1 and 3 months postoperatively.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!