Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: ADVANCE Medial-Pivot (MP) (Wright Medical Technology, Arlington, TN, USA) total knee arthroplasty (TKA) was developed to replicate normal tibiofemoral knee joint kinematics, allowing medial-pivot knee motion. The design concept of the prosthesis is unique; therefore, the influence on the patellofemoral knee joint remains unclear at present. The purpose of this study was to determine the in vivo patellofemoral kinematics with ADVANCE MP TKA and compare them with the pre-operative conditions.
Methods: ADVANCE MP TKA was performed in ten subjects with osteoarthritis (OA). At before and one month after surgery, lateral radiographs with weight-bearing at maximum extension, 30, 60 and 90° were taken, and patella flexion angle (PF), tibiopatellar angle (TP) and estimated patellofemoral contact point (PC) were evaluated, according to a previously reported method.
Results: In PF and TP, there was no statistically significant change between pre-operative and postoperative values. Pre-operative PC reached its peak at 90°; however, its peak was at 60° at one month after surgery. Postoperative PC at maximum extension was significantly higher compared to before surgery.
Conclusions: The results in this study indicated that ADVANCE MP TKA changed patellofemoral joint kinematics compared to before surgery. Early postoperative evaluation is the limitation of this study; however, we consider that the results in this study might be one of the keys to resolving the kinematic features of this prosthesis, helping clinicians to comprehend this prosthesis.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3460083 | PMC |
http://dx.doi.org/10.1007/s00264-012-1634-2 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!