Background: We tested the hypothesis that the shape of the shock waveform affects not only the defibrillation threshold but also the amount of cardiac damage.

Methods And Results: Defibrillation thresholds were determined for 11 waveforms-3 ascending-ramp waveforms, 3 descending-ramp waveforms, 3 rectilinear first-phase biphasic waveforms, a Gurvich waveform, and a truncated exponential biphasic waveform-in 6 pigs with electrodes in the right ventricular apex and superior vena cava. The ascending, descending, and rectilinear waveforms had 4-, 8-, and 16-millisecond first phases and a 3.5-millisecond rectilinear second phase that was half the voltage of the first phase. The exponential biphasic waveform had a 60% first-phase and a 50% second-phase tilt. In a second study, we attempted to defibrillate after 10 seconds of ventricular fibrillation with a single ≈30-J shock (6 pigs successfully defibrillated with 8-millisecond ascending, 8-millisecond rectilinear, and truncated exponential biphasic waveforms). Troponin I blood levels were determined before and 2 to 10 hours after the shock. The lowest-energy defibrillation threshold was for the 8-milliseconds ascending ramp (14.6±7.3 J [mean±SD]), which was significantly less than for the truncated exponential (19.6±6.3 J). Six hours after shock, troponin I was significantly less for the ascending-ramp waveform (0.80±0.54 ng/mL) than for the truncated exponential (1.92±0.47 ng/mL) or the rectilinear waveform (1.17±0.45 ng/mL).

Conclusions: The ascending ramp has a significantly lower defibrillation threshold and at ≈30 J causes 58% less troponin I release than the truncated exponential biphasic shock. Therefore, the shock waveform affects both the defibrillation threshold and the amount of cardiac damage.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3968773PMC
http://dx.doi.org/10.1161/CIRCULATIONAHA.112.109777DOI Listing

Publication Analysis

Top Keywords

truncated exponential
24
defibrillation threshold
20
exponential biphasic
20
biphasic waveform
12
waveform
8
lower defibrillation
8
shock waveform
8
waveform defibrillation
8
threshold amount
8
amount cardiac
8

Similar Publications

Fast Proxy Centers for the Jeffreys Centroid: The Jeffreys-Fisher-Rao Center and the Gauss-Bregman Inductive Center.

Entropy (Basel)

November 2024

Sony Computer Science Laboratories, Tokyo 141-0022, Japan.

The symmetric Kullback-Leibler centroid, also called the Jeffreys centroid, of a set of mutually absolutely continuous probability distributions on a measure space provides a notion of centrality which has proven useful in many tasks, including information retrieval, information fusion, and clustering. However, the Jeffreys centroid is not available in closed form for sets of categorical or multivariate normal distributions, two widely used statistical models, and thus needs to be approximated numerically in practice. In this paper, we first propose the new Jeffreys-Fisher-Rao center defined as the Fisher-Rao midpoint of the sided Kullback-Leibler centroids as a plug-in replacement of the Jeffreys centroid.

View Article and Find Full Text PDF

Direct Givens rotation method based on error back-propagation algorithm for self-consistent field solution.

J Chem Phys

January 2025

Department of Chemistry and Biochemistry, School of Advanced Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo 169-8555, Japan.

The self-consistent field (SCF) procedure is the standard technique for solving the Hartree-Fock and Kohn-Sham density functional theory calculations, while convergence is not theoretically guaranteed. Direct minimization methods, such as the augmented Lagrangian method (ALM) and second-order SCF (SOSCF), obtain the SCF solution by minimizing the Lagrangian with the gradient. In SOSCF, molecular orbitals are optimized by truncating the Taylor expansion of a unitary matrix represented in exponential form to ensure the orthonormality condition.

View Article and Find Full Text PDF

Evaluation of current patient-based real-time quality control in clinical chemistry testing.

Clin Chim Acta

December 2024

Department of Clinical Chemistry, Department of Biochemistry, University of Health Sciences, Bakirköy Dr. Sadi Konuk Training and Research Hospital, İstanbul, Türkiye.

Introduction: To perform simulation studies on patient-based real-time quality control (PBRTQC) for aspartate aminotransferase (AST), iron (Fe), potassium (K), and thyrotropin (thyroid stimulating hormone, TSH) analytes, focusing on optimizing systematic error detection while minimizing data loss.

Methods: Clinical laboratory data for the four analytes were analyzed using various truncation methods. Among these methods, truncation limits corresponding to fixed percentiles (e.

View Article and Find Full Text PDF

Control charts are commonly used for process monitoring under the assumption that the variable of interest follows a normal distribution. However, this assumption is frequently violated in real-world applications. In this study, we develop an adaptive control chart based on the exponentially weighted moving average (EWMA) statistic to monitor irregular variations in the mean of the Truncated Transmuted Burr-II (TTB-II) distribution, employing Hastings approximation for normalization.

View Article and Find Full Text PDF

Background: The activation of hepatic stellate cells play a pivotal role in the pathogenesis of hepatic fibrosis. However, the current lack of specifically identified targets on these cells poses a significant challenge in developing targeted delivery tools for effective anti-hepatic fibrosis therapeutics in clinical practice.

Methods: Cell-systematic evolution of ligands by exponential enrichment method was conducted on HSC-T6 cell line to screen out activated hepatic stellate cell-specific aptamers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!