Feasibility study of automated overnight closed-loop glucose control under MD-logic artificial pancreas in patients with type 1 diabetes: the DREAM Project.

Diabetes Technol Ther

Jesse Z. and Sara Lea Shafer Institute for Endocrinology and Diabetes, The National Center for Childhood Diabetes, Schneider Children's Medical Center of Israel, 14 Kaplan St., Petah Tikva, Israel .

Published: August 2012

Background: Artificial pancreas systems may offer a potential major impact on the normalization of metabolic control and preventing hypoglycemic events. This study aims to establish near-normal overnight glucose control and reduce the risk of nocturnal hypoglycemia using the MD-Logic Artificial Pancreas (MDLAP), an algorithm that was developed by our research group. This inpatient feasibility study is the first step towards implementing an overnight closed-loop MDLAP system at the patient's home.

Subjects And Methods: Seven patients with type 1 diabetes (three adolescents and four adults; mean±SD age, 20.6±4.7 years; duration of diabetes, 9.6±2.6 years; body mass index, 24.3±3.9 kg/m(2); and glycated hemoglobin, 7.8±0.8%) participated in a total of 14 closed-loop overnight sessions. Each participant underwent two closed-loop inpatient sessions starting at dinner alone and at dinner following exercise. The closed-loop inpatient sessions were compared with data derived from nights spent at home with an open-loop system in a similar scenario to the study protocol.

Results: The mean percentage of time spent in the near normal glucose range of 63-140 mg/dL was 83±16%, and the median (interquartile range) was 85% (78-92%) for the overnight closed-loop sessions compared with 34±31% and 27% (6-57%) in the homecare open-loop setting, respectively. During the overnight closed-loop sessions at dinner alone 92±9% of the sensor values ranged within target range, compared with 73±19% for the sessions following exercise (P=0.03). No hypoglycemic (<63 mg/dL) events occurred during the closed-loop sessions.

Conclusion: Closed-loop insulin delivery under MDLAP is a feasible and safe solution to control overnight glycemia.

Download full-text PDF

Source
http://dx.doi.org/10.1089/dia.2012.0004DOI Listing

Publication Analysis

Top Keywords

overnight closed-loop
16
artificial pancreas
12
feasibility study
8
glucose control
8
md-logic artificial
8
patients type
8
type diabetes
8
closed-loop inpatient
8
inpatient sessions
8
sessions compared
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!