Aims: To investigate the behaviour of the implantable cardioverter defibrillator (ICD) function during actual radiotherapy sessions.

Methods And Results: Fifteen patients with an ICD underwent 17 radiation treatments for cancer [cumulative dose to the tumour was between 16 Gray (Gy) and 70 Gy; photon beams with maximum energies between 6 megaelectronvolt (MeV) and 18 MeV were employed]. During every session, the ICD was programmed to a monitoring mode to prevent inappropriate therapy delivery. Afterwards, the ICDs were interrogated to ensure proper function. Calculated radiation dose at the ICD site was <1 Gy in all patients. In 5 out of 17 radiation treatments (29%) the ICDs showed 6 malfunctions (35%). We noticed four disturbances in the memory data or device resets during radiation treatment and one case of inappropriate ventricular fibrillation detection due to external noise. In one case a late device data error was observed. All malfunctions occurred at 10 and 18 MeV beam energies.

Conclusion: Despite the fact that all recommended precautions were taken to minimize the damage to the ICDs during radiotherapy and the calculated dose to the ICDs was <1 Gy, in 29% of the treatments a malfunction occurred. We observed a possible correlation between the beam energy and the malfunctions. This correlation may be due to an interaction between neutrons produced in the head of the linear accelerator at beam energies ≥10 MeV, and boron-10 which is present in the integrated circuit.

Download full-text PDF

Source
http://dx.doi.org/10.1093/europace/eus197DOI Listing

Publication Analysis

Top Keywords

implantable cardioverter
8
cardioverter defibrillator
8
high incidence
4
incidence implantable
4
defibrillator malfunctions
4
malfunctions radiation
4
radiation therapy
4
therapy neutrons
4
neutrons probable
4
probable soft
4

Similar Publications

Introduction: The role of the sympathetic nervous system in the initiation and continuation of ventricular tachyarrhythmias (VTA) is well established. However, whether CSD reduces implantable cardioverter-defibrillator (ICD) shocks and recurrent VTA is still uncertain.

Methods: A comprehensive literature search was performed at Medline and Embase until March 2023.

View Article and Find Full Text PDF

Background: Guideline-recommended strategies to interrupt chronic anticoagulation with warfarin or direct oral anticoagulants (DOAC) during the perioperative period of cardiac implantable electronic device (CIED) surgery differ worldwide. There is uncertainty concerning the benefits and harms of interrupted and uninterrupted anticoagulation in patients undergoing CIED surgery.

Objectives: To assess the benefits and harms of interrupted anticoagulation (IAC) with either warfarin or DOAC in the perioperative period of CIED surgery versus uninterrupted anticoagulation (UAC), with or without heparin bridging, during an equivalent time frame, for CIED surgery.

View Article and Find Full Text PDF

Sudden cardiac death is a leading cause of mortality in children with hypertrophic cardiomyopathy (HCM). The PRecIsion Medicine in CardiomYopathy consortium developed a validated tool (PRIMaCY) for sudden cardiac death risk prediction to help with implantable cardioverter defibrillator shared decision-making, as recommended by clinical practice guidelines. The mplemeting a udden Cardiac Dath isk Assessment ool in hildhood (INSERT-HCM) study aims to implement PRIMaCY into electronic health records (EHRs) and assess implementation determinants and outcomes.

View Article and Find Full Text PDF

Background: The evidence in the primary prevention of sudden cardiac death (SCD) by using implantable cardioverter defibrillators (ICD) in patients with ischemic cardiomyopathy (ICM) is well-established but remains controversial for those with nonischemic cardiomyopathy (NICM). This study evaluates whether prognostic differences exist between ICM and NICM patients after ICD implantation.

Methods: Patients initially undergoing ICD implantation as primary SCD prevention were retrospectively enrolled from January 2017 to May 2023.

View Article and Find Full Text PDF

Inappropriate Shocks in Brugada Syndrome Patients With a Subcutaneous Implantable Cardioverter Defibrillator.

Pacing Clin Electrophysiol

January 2025

Department of Cardiovascular Medicine, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Sapporo, Japan.

Background: This study aimed to compare inappropriate shock (IAS) rates between subcutaneous implantable cardioverter-defibrillator (S-ICD) and transvenous ICD (TV-ICD) in Brugada syndrome (BrS) patients and identify risk factors for IAS in S-ICD use.

Methods: We enrolled consecutive patients with BrS who underwent ICD implantation between 2013 and 2023. Data on clinical characteristics, S-ICD screening test data, and IAS occurrence were retrospectively analyzed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!