Influenza infections are associated with thousands of hospital admissions and deaths each year. Rapid detection of influenza is important for prompt initiation of antiviral therapy and appropriate patient triage. In this study the Cepheid Xpert Flu assay was compared with two rapid antigen tests, BinaxNOW Influenza A & B and BD Directigen EZ Flu A+B, as well as direct fluorescent antibody testing for the rapid detection of influenza A and B. Using real-time, hydrolysis probe-based, reverse transcriptase PCR as the reference method, influenza A sensitivity was 97.3% for Xpert Flu, 95.9% for direct fluorescent antibody testing, 62.2% for BinaxNOW, and 71.6% for BD Directigen. Influenza B sensitivity was 100% for Xpert Flu and direct fluorescent antibody testing, 54.5% for BinaxNOW, and 48.5% for BD Directigen. Specificity for influenza A was 100% for Xpert Flu, BinaxNOW, and BD Directigen, and 99.2% for direct fluorescent antibody testing. All methods demonstrated 100% specificity for influenza B. These findings support the use of the Xpert Flu assay in settings requiring urgent diagnosis of influenza A and B.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4075941 | PMC |
http://dx.doi.org/10.1016/j.jviromet.2012.07.023 | DOI Listing |
Background: Early confirmation of infections with influenza virus and/or respiratory syncytial virus (RSV) is beneficial for prompt treatment and outbreak management. This study aimed to assess the Cepheid Xpert Xpress Flu/ RSV assay in Central China, using Sanger sequencing as the reference method.
Methods: Nasopharyngeal swab (NP) samples from pediatric and adult patients with influenza-like illnesses were collected by the Hubei Province Disease Control and Prevention Center.
PLoS One
November 2024
Cepheid, Sunnyvale, California, United States of America.
While molecular testing is recommended for symptomatic patients suspected of having coronavirus disease 2019 (COVID-19), limited data are available examining real-world use of tests for severe acute respiratory syndrome coronavirus (SARS-CoV-2) and the impact of SARS-CoV-2 testing on patient outcomes. In this retrospective cohort study using de-identified administrative claims data in the Optum Labs Data Warehouse, we identified 2 groups of patients with ≥1 outpatient claims with a procedure code for SARS-CoV-2 testing between January 2021 and September 2022. Group 1 had ≥1 claims with CPT code 0240U or 0241U ("Xpert Xpress") (N = 51,602); Group 2 had ≥1 claims for laboratory-based molecular testing (N = 317,192).
View Article and Find Full Text PDFDiagnostics (Basel)
October 2024
Division of Clinical Pathology, Department of Pathology, National Defense Medical Center, Tri-Service General Hospital, Taipei 114, Taiwan.
J Virol Methods
January 2025
Ndlovu Research Centre, Ndlovu Laboratories, Elandsdoorn, Dennilton, Limpopo, South Africa. Electronic address:
Microbiol Spectr
August 2024
Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
Unlabelled: Seasonal increase of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza virus A/B (Flu A/B), and respiratory syncytial virus (RSV) require rapid diagnostic test methods for the management of respiratory tract infections. In this study, we compared the diagnostic accuracy of Savanna RVP4 (RVP4, QuidelOrtho) with Xpert Xpress Plus SARS-CoV-2/Flu/RSV (Xpert, Cepheid). Nasopharyngeal swabs from patients treated at a tertiary care hospital (Germany) were tested for SARS-CoV-2, Flu A/B, and RSV by RVP4 to assess diagnostic accuracy (reference standard: Xpert).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!