Objective: To establish an ion chromatography (IC) method for determination of ammonia in air of workplace.
Methods: Ammonia in workplace air was collected in silica gel tube, desorbed with 10 mmol/L methanesulfonic acid by ultrasonic for 10 min, determined by IC.
Results: The linearity range was 0.02-1.00 microg/ml. The linear equation was Y = 12041X-187 (r = 0.9997). The limit of quantification was 0.13 mg/m3 (the air volume was 1.5 L). Collection efficiency was 100%. Extraction efficiency was 99%. The relative standard deviation was 4.2%-6.3%. The penetration capacity was more than 264 microg. Sample could be stored for 14 days at least by ambient storage.
Conclusion: This method is convenient, applicable and sensitive, suitable to determinate ammonia in air of workplace.