Context: Proper diagnosis and therapy of fibrinogen deficiency requires high-quality fibrinogen assays.

Objective: To assess the interlaboratory bias, precision, and grading of fibrinogen assays used by laboratories participating in the United States College of American Pathologists proficiency testing program in coagulation.

Design: Two identical vials of normal plasma were sent to more than 3500 laboratories. Participants measured fibrinogen levels using local methods.

Results: Fifty different fibrinogen methods were evaluated. All-method bias was 8.3% (range of method-specific biases, 0.0%-27.0%) and all-method coefficient of variation was 7.7% (range of method-specific coefficients of variation, 0.7%-25.8%). After controlling for reagent/instrument type, mean fibrinogen levels were 11.6% higher for prothrombin time-based reagents compared to Clauss (P < .001), and coefficient of variation was 46% lower for mechanical endpoint instruments compared to photo-optical. Most testing events (97.4%) could be reliably graded as pass or fail using a target range of ±20% from the method mean (total pass rate, 98.8%). Total fail rate was 3.0-fold lower for mechanical instruments compared to photo-optical (0.5% versus 1.5%, P  =  .001). Nonetheless many photo-optical methods had very high precision and very low fail rates.

Conclusions: Fibrinogen assays showed highly variable methodology and performance characteristics. Bias, precision, and grading were affected by the type of reagent or instrument used.

Download full-text PDF

Source
http://dx.doi.org/10.5858/arpa.2011-0322-OADOI Listing

Publication Analysis

Top Keywords

fibrinogen assays
12
fibrinogen
8
normal plasma
8
college american
8
american pathologists
8
pathologists proficiency
8
proficiency testing
8
testing program
8
bias precision
8
precision grading
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!