Objective: To determine the effect of modifying the GlideScope (GVL) blade on the intubation time.

Methods: This prospective study was conducted at the Department of Anesthesia, King Abdulaziz University Hospital, Jeddah, Saudi Arabia between June 2011 and October 2011. Sixty patients requiring endotracheal tube (ETT) intubation for elective surgery in whom airway was anticipated normal were randomly allocated to one of 2 groups. Group M (n=30): intubated via a modified GVL blade in which a tube conduit along the side of the GVL blade was created to allow the passage of ETT through the cords. Group C (n=30): intubated with the conventional GVL blade and rigid intubating stylet.

Results: Time to successful tracheal intubation (TTI) was 39.6+/-2.1 seconds in Group M versus 66.4+/-8.3 seconds in Group C (p=0.0001), tracheal intubation was deemed more easily in Group M than in Group C (VAS 2+/-1 versus 6+/-1, p=0.0001), and all patients in Group M were successfully intubated on the first attempt when compared with 90% in Group C (p=0.009).

Conclusion: The addition of a conduit to the GVL blade made the passage of the ETT easier and TTI shorter without increasing adverse events or intubation failure.

Download full-text PDF

Source

Publication Analysis

Top Keywords

gvl blade
20
tracheal intubation
12
group
8
group n=30
8
n=30 intubated
8
passage ett
8
seconds group
8
blade
6
intubation
6
gvl
5

Similar Publications

Background: Postoperative sore throat (POST) is a common complaint after general anesthesia. POST is defined by the presence of a self-reported foreign body sensation or a painful sensation in the throat after general anesthesia. This condition may affect recovery and patient satisfaction and is associated with many factors, including intubation tools.

View Article and Find Full Text PDF

Background: The GlideScope® videolaryngoscope (GVL) is widely used in patients with difficult airways and provides a good glottic view. However, the acute angle of the blade can make insertion and advancement of an endotracheal tube (ETT) more difficult than direct laryngoscopy, and the use of a stylet is recommended. This randomized controlled trial compared Parker Flex-It™ stylet (PFS) with GlideRite® rigid stylet (GRS) to facilitate intubation with the GVL in simulated difficult intubations.

View Article and Find Full Text PDF

Comparison of the strength of various disposable videolaryngoscope blades.

Can J Anaesth

November 2021

Department of Emergency Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea.

Purpose: Breaking of disposable blades during emergency endotracheal intubation has been reported. Breakage can cause serious injury and foreign body ingestion. We aimed to measure and analyze the strength characteristics of different disposable videolaryngoscope blades with the application of an upward-lifting force.

View Article and Find Full Text PDF

Background And Aims: Glidescope®videolaryngoscope (GVL) is a video intubation system with 60° angle blade that provides excellent laryngeal view, does not require alignment of oral, pharyngeal, and laryngeal axes for visualisation of glottis, thus causing less stimulation of orolaryngopharynx. The aim of this study was to compare haemodynamic responses (blood pressure and heart rate) and airway morbidity using the Macintosh direct laryngoscope (MDL) and the Glidescope®videolaryngoscope (GVL) in hypertensive patients.

Methods: Fifty patients with hypertension controlled on antihypertensive medications scheduled for elective surgery under general anaesthesia were randomly assigned to group GVL ( = 25) or group MDL ( = 25).

View Article and Find Full Text PDF

A randomized controlled trial comparing nociception level (NOL) index, blood pressure, and heart rate responses to direct laryngoscopy versus videolaryngoscopy for intubation: the NOLint project.

Can J Anaesth

June 2021

Department of Anesthesiology and Pain Medicine of University of Montreal, and the Department of Anesthesiology and Pain Medicine of Maisonneuve Rosemont Hospital, CIUSSS de l'Est de l'Ile de Montréal, University of Montreal, 5415, Boulevard de l'Assomption, Montréal, QC, H1T 2M4, Canada.

Purpose: The effect of direct laryngoscopy using a Macintosh blade (MAC) vs GlideScope™ videolaryngoscopy using a Spectrum LoPro blade (GVL) on nociceptive stimulation has not been quantitatively studied. This study used the new nociception level (NOL) index to compare the nociceptive response induced by GVL or MAC during laryngoscopy with or without intubation.

Methods: Patients underwent two laryngoscopies at four-minute intervals (L1, L2), one with GVL and the other with MAC (first randomization).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!