Background: Albuminuria is an important sign of chronic kidney disease and is detected routinely by measurement of urinary albumin-creatinine ratio (ACR). A Siemens CLINITEK test designed for use at the point of care is available that can semiquantitatively measure ACR.

Study Design: Diagnostic accuracy study evaluating a urinary ACR point-of-care test.

Setting & Participants: The semiquantitative ACR test was evaluated at the point of care in a representative primary care population (those with or at increased risk of chronic kidney disease) of 642 patients under standard operational conditions and compared with the reference standard of ACR measurement in the clinical laboratory.

Index Test: The point-of-care CLINITEK semiquantitative ACR test. This test uses dye-binding and catalytic assays for albumin and creatinine, respectively, on a Microalbumin 9 strip, which is read by the CLINITEK Status Analyzer, and ACR is calculated automatically.

Reference Test: Laboratory measurement of albumin and creatinine on an Abbott Architect analyzer by immunoturbidimetric and enzymatic assays, respectively, and calculation of ACR.

Results: The prevalence of albuminuria (laboratory ACR≥30 mg/g) in the study population was 20.2%. Sensitivity and specificity of the point-of-care test for detecting albuminuria were 83.2% and 80.0%, respectively. Positive and negative predictive values were 51.2% and 95.0%, respectively; positive and negative likelihood ratios were 4.16 and 0.21, respectively. Twenty-three (3.6%) samples measured at the point of care were not analyzed in the central laboratory for a variety of reasons, including laboratory reception data entry errors.

Limitations: Our sensitivity calculation is accurate to an approximately 8% CI.

Conclusions: The instrument-read reagent strip test was a poor rule-in test for albuminuria at the point of care, as evidenced by the low positive predictive value, but was a reasonable rule-out test. Observed sensitivity was lower than reported in earlier laboratory-based studies. This decreased diagnostic accuracy needs to be balanced against the potential advantages of a point-of-care testing approach.

Download full-text PDF

Source
http://dx.doi.org/10.1053/j.ajkd.2012.05.009DOI Listing

Publication Analysis

Top Keywords

point care
16
diagnostic accuracy
12
test
11
albumin-creatinine ratio
8
point-of-care test
8
primary care
8
chronic kidney
8
kidney disease
8
semiquantitative acr
8
acr test
8

Similar Publications

Point-of-care ultrasound in the diagnosis of hepatic gas gangrene.

J Ultrasound

January 2025

Argentinian Critical Care Ultrasonography Association (ASARUC), Buenos Aires, Argentina.

Hepatic gas gangrene (HGG) is a rare but life-threatening condition typically caused by anaerobic bacteria such as Clostridium perfringens, though Gram-negative bacteria like Escherichia coli and Klebsiella species have also been implicated. Traditionally diagnosed via computed tomography (CT), point-of-care ultrasound (POCUS) has emerged as a valuable tool in critical care settings for its non-invasive, bedside utility. We report the case of a 51-year-old female with choledochal syndrome secondary to cholangiocarcinoma who developed HGG following left extended hepatectomy and biliary reconstruction.

View Article and Find Full Text PDF

Introduction: To target psychological support to cancer patients most in need of support, screening for psychological distress has been advocated and, in some settings, also implemented. Still, no prior studies have examined the appropriate 'dosage' and whether screening for distress before cancer treatment may be sufficient or if further screenings during treatment are necessary. We examined the development in symptom trajectories for breast cancer patients with low distress before surgery and explored potential risk factors for developing burdensome symptoms at a later point in time.

View Article and Find Full Text PDF

Background: Point-of-care ultrasound (POCUS) can be used in a variety of clinical settings and is a safe and powerful tool for ultrasound-trained healthcare providers, such as physicians and nurses; however, the effectiveness of ultrasound education for nursing students remains unclear. This prospective cohort study aimed to examine the sustained educational impact of bladder ultrasound simulation among nursing students.

Methods: To determine whether bladder POCUS simulation exercises sustainably improve the clinical proficiency regarding ultrasound examinations among nursing students, evaluations were conducted before and after the exercise and were compared with those after the 1-month follow-up exercise.

View Article and Find Full Text PDF

The optimal duration of on-scene cardiopulmonary resuscitation (CPR) for out-of-hospital cardiac arrest (OHCA) patients remains uncertain. Determining this critical time period requires outweighing the potential risks associated with intra-arrest transport while minimizing delays in accessing definitive hospital-based treatments. This study evaluated the association between on-scene CPR duration and 30-day neurologically favorable survival based on the transport time interval (TTI) in patients with OHCA.

View Article and Find Full Text PDF

The accessibility of CAR-T cells in centralized production models faces significant challenges, primarily stemming from logistical complexities and prohibitive costs. However, European Regulation EC No. 1394/2007 introduced a pivotal provision known as the hospital exemption.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!