A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Low-dose CT of the lung: potential value of iterative reconstructions. | LitMetric

Low-dose CT of the lung: potential value of iterative reconstructions.

Eur Radiol

Institute for Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091, Zurich, Switzerland.

Published: December 2012

Objectives: To prospectively assess the impact of sinogram-affirmed iterative reconstruction (SAFIRE) on image quality of nonenhanced low-dose lung CT as compared to filtered back projection (FBP).

Methods: Nonenhanced low-dose chest CT (tube current-time product: 30 mAs) was performed on 30 patients at 100 kVp and on 30 patients at 80 kVp. Images were reconstructed with FBP and SAFIRE. Two blinded, independent readers measured image noise; two readers assessed image quality of normal anatomic lung structures on a five-point scale. Radiation dose parameters were recorded.

Results: Image noise in datasets reconstructed with FBP (57.4 ± 15.9) was significantly higher than with SAFIRE (31.7 ± 9.8, P < 0.001). Image quality was significantly superior with SAFIRE than with FBP (P < 0.01), without significant difference between FBP at 100 kVp and SAFIRE at 80 kVp (P = 0.68). Diagnostic image quality was present with FBP in 96% of images at 100 kVp and 88% at 80 kVp, and with SAFIRE in 100% at 100 kVp and 98% at 80 kVp. There were significantly more datasets with diagnostic image quality with SAFIRE than with FBP (P < 0.01). Mean CTDI(vol) and effective doses were 1.5 ± 0.7 mGy·cm and 0.7 ± 0.2 mSv at 100 kVp, and 1.4 ± 2.8 mGy · cm and 0.5 ± 0.2 mSv at 80 kVp (P < 0.001, both).

Conclusions: Use of SAFIRE in low-dose lung CT reduces noise, improves image quality, and renders more studies diagnostic as compared to FBP.

Key Points: Low-dose computed tomography is an important thoracic investigation tool. Radiation dose can be less than 1 mSv with iterative reconstructions. Iterative reconstructions render more low-dose lung CTs diagnostic compared to conventional reconstructions.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-012-2524-0DOI Listing

Publication Analysis

Top Keywords

image quality
24
100 kvp
20
low-dose lung
16
iterative reconstructions
12
kvp
10
safire
8
image
8
nonenhanced low-dose
8
reconstructed fbp
8
image noise
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!