Purpose: Pulmonary edema (PE) is associated with fluid accumulation in the lungs. Device-based impedance measurements have been used to detect fluid overload prior to hospitalization. However, studies have reported a high false positive rate (FPR). The objective of this study was to develop and test a new multivector impedance-based algorithm that reliably tracks PE clinical events.
Methods: We enrolled patients with implanted CRT-Ds in 23 US centers within 2 weeks of device implant. Six-vector impedance data was collected automatically by the CRT-Ds every 30 min during emergency department visits/hospitalizations and every 2 h at all other times. Detection algorithms for cardiac resynchronization therapy defibrillator (CRT-D) and implantable cardiac defibrillator (ICD) devices were developed using those impedance vectors that would be available in corresponding devices and retrospectively evaluated.
Results: There were 75 patients (69 % male), mean age 66 ± 12 years, with a LVEF of 23 ± 6 % and QRS of 149 ± 25 ms. Twenty-one major clinical events occurred over 8.2 ± 2.6 months of follow-up time. CRT-D vector combinations resulted in a sensitivity of 71.4 % (95 % confidence interval 47.8-88.7) and a FPR of 0.56 (0.30-0.94) false positives per patient-year (FPs/pt-yr); ICD vector combinations resulted in a sensitivity of 61.9 % (38.4-81.9) and a FPR of 0.63 (0.36-0.90) FPs/pt-yr. In comparison, the single-vector RVCoil-Can implementation of this algorithm resulted in a sensitivity of 57.1 % (34.0-78.2) and a FPR of 0.74 (0.44-1.12) FPs/pt-yr.
Conclusions: This multivector impedance algorithm was effective in tracking PE clinical events in this patient population. Additional studies are needed to prospectively evaluate the performance of this algorithm in a larger population.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10840-012-9693-2 | DOI Listing |
Int J Cardiol Heart Vasc
September 2015
The Department of Internal Medicine, The Ohio State University, United States.
Background: A significant proportion of patients admitted for acute decompensated heart failure (ADHF) that undergo volume reduction therapy are discharged with unchanged or increased bodyweight suggesting that the endpoints for these therapies are not optimally defined. We aimed to identify vectors that can help monitor changes in intravascular fluid volume, that in turn may more accurately guide volume reduction therapy.
Methods: Data from six different impedance vectors and corresponding changes in intravascular volume derived from changes in hematocrit were obtained from 132 clinical congestion events in 56 unique patients enrolled in a multisite trial of early detection of clinical congestion events (DEFEAT PE).
J Cardiovasc Med (Hagerstown)
October 2013
aDivision of Cardiology, Department of Internal Medicine, University of Rome 'Tor Vergata',, Rome, Italy bElectrophysiology, Pacemaker and ICD Unit, Hygeia Hospital, Athens, Greece cNational Research Council (CNR), Institute of Clinical Physiology, Section of Rome, Rome, Italy.
Objectives: We investigated the performance of a new intrathoracic multivector impedance monitoring system for the prediction of heart failure events in consecutive device-implanted patients.
Methods: Eighty heart failure patients implanted with biventricular defibrillators with multivector impedance monitoring capability were prospectively enrolled. Clinical heart failure status and impedance data were assessed during follow-up and if patients presented with an alert or heart failure deterioration.
J Interv Card Electrophysiol
November 2012
Division of Cardiovascular Medicine, College of Medicine, College of Public Health, The Ohio State University, Columbus, OH, USA.
Purpose: Pulmonary edema (PE) is associated with fluid accumulation in the lungs. Device-based impedance measurements have been used to detect fluid overload prior to hospitalization. However, studies have reported a high false positive rate (FPR).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!