AI Article Synopsis

  • The study aimed to evaluate the accuracy and consistency of central corneal thickness (CCT) measurements using three different methods: swept-source anterior segment optical coherence tomography (AS-OCT), spectral-domain retinal OCT, and ultrasound pachymetry (USP).
  • The analysis involved 50 healthy volunteers and utilized statistical methods to compare CCT measurements and assess repeatability, revealing that AS-OCT provided the most consistent results.
  • Although CCT measurements from all three methods showed strong correlations, they were not identical, highlighting AS-OCT’s excellent repeatability in producing pachymetric maps.

Article Abstract

Purpose: To assess accuracy and repeatability of central corneal thickness (CCT) measurements obtained by swept-source anterior segment optical coherence tomography (AS-OCT), spectral-domain retinal OCT with corneal module and ultrasound pachymetry (USP), and to assess repeatability of pachymetric mapping with AS-OCT.

Methods: 50 healthy volunteers were recruited. A single, experienced operator analysed the right eye of each participant twice in the same session with AS-OCT ('corneal map' routine), retinal OCT and USP. CCT measurements were compared using repeated-measures analysis of variance, Bonferroni test, Pearson correlation and Bland-Altman plots. Repeatability of thickness maps and CCT measurements were assessed using Alpha of Cronbach, intraclass correlation coefficient (ICC) and coefficient of repeatability.

Results: Mean CCT±SD was 540±28.9 μm for AS-OCT, 544±29.5 μm for retinal OCT and 549.3±31.7 μm for USP; the differences were statistically significant (p<0.01). CCT measurements obtained with the three instruments were highly correlated: r was 0.965 for AS-OCT/USP, 0.962 for retinal OCT/USP and 0.984 for AS-OCT/retinal OCT comparison. The repeatability of CCT measurements was higher for AS-OCT than for the other devices (p<0.001). Repeatability of pachymetric maps was excellent (ICC=0.999).

Conclusions: Pachymetric maps by swept-source AS-OCT showed excellent repeatability. CCT measurements obtained by AS-OCT, USP and retinal OCT were highly correlated although not identical.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1755-3768.2012.02453.xDOI Listing

Publication Analysis

Top Keywords

cct measurements
12
retinal oct
12
corneal thickness
8
swept-source anterior
8
anterior segment
8
segment optical
8
optical coherence
8
coherence tomography
8
thickness mapping
4
mapping swept-source
4

Similar Publications

Biomarkers.

Alzheimers Dement

December 2024

The First Affiliated Hospital of Chongqing Medical University, Chongqing, Chongqing, China.

Background: The regulatory role of Trimethylamine-N-oxide (TMAO) for cognition from the perspective of microbiota-gut-brain (MGB) axis in AD remains unclear.

Method: In clinical cohort study for effects of 24-week computerized cognitive training (CCT), registered on clinicaltrials.gov (NCT06094452), plasma TMAO levels were quantified using ELISA in MCI (n=39) and mild AD patients (n=35).

View Article and Find Full Text PDF

Purpose: To investigate long-term corneal biomechanical changes in thin corneas after small incision lenticule extraction (SMILE).

Methods: Patients with indications for SMILE were enrolled in this study between November 2017 and March 2018. Patients were matched for age, spherical diopter, cylinder, spherical equivalent (SE), and lenticule thickness (LT), and then categorized into the thin cornea group (preoperative thinnest central corneal thickness [CCT] of 500 µm or less, 32 eyes) or normal cornea group (CCT of greater than 500 µm, 32 eyes).

View Article and Find Full Text PDF

Layer-by-Layer (LbL) self-assembly encapsulation is a promising technology for the protection and delivery of lactic acid bacteria. However, laboratory-scale encapsulation is often time-consuming, involves intensive protocols tailored for small-scale operations, requires substantial amounts of energy and water, and results in a low yield of encapsulated biomass. Scaling-up this process to a bench-bioreactor scale is not simply a matter of increasing culture volume as different key parameters (not particularly relevant at lab scale) become critical, including biomass production, the number of polymer layers, and the biomass-to-polymer mass ratio.

View Article and Find Full Text PDF

Purpose: the purpose of this study was to evaluate the safety and usability of the ATLAS 2030 in children with Cerebral Palsy (CP) and Spinal Muscular Atrophy (SMA).

Materials And Methods: the sample consisted of six children, three with CP and three with SMA, who received eight sessions of robot-assisted gait therapy. Safety was measured by the presence of adverse events.

View Article and Find Full Text PDF

Gillies ex Hooker et Arnott (Oxalidaceae): Chemical Analysis, Biological In Vitro and In Vivo Properties and Behavioral Effects.

Antioxidants (Basel)

December 2024

Consejo Nacional de Investigaciones Científicas y Técnicas (CONICET), Ciudad Autónoma de Buenos Aires C1425FQB, Argentina.

In this work, a decoction (DOe) and a methanolic global extract (MGEOe), obtained with the aerial parts of Gillies ex Hooker et Arnott (Oxalidaceae), were evaluated. The high-resolution liquid chromatography in conjunction with electrospray ionization quadrupole time-of-flight mass spectrometry (UHPLC-ESI-QTOF-MS) analysis showed forty compounds in MGEOe and twenty-nine in DOe, including flavones, C-glycosyl flavones, isoflavones, fatty acids, terpenes, phenolic acids, and sterols. The antioxidant properties were evaluated by DPPH, TEAC, FRAP, and ILP assays.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!