Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery.

Pediatr Radiol

Department of X-ray Physics, Region of Northern Jutland, Broenderslev, Denmark.

Published: August 2012

Background: During surgery for spinal deformities, accurate placement of pedicle screws may be guided by intraoperative cone-beam flat-detector CT.

Objective: The purpose of this study was to identify appropriate paediatric imaging protocols aiming to reduce the radiation dose in line with the ALARA principle.

Materials And Methods: Using O-arm® (Medtronic, Inc.), three paediatric phantoms were employed to measure CTDI(w) doses with default and lowered exposure settings. Images from 126 scans were evaluated by two spinal surgeons and scores were compared (Kappa statistics). Effective doses were calculated. The recommended new low-dose 3-D spine protocols were then used in 15 children.

Results: The lowest acceptable exposure as judged by image quality for intraoperative use was 70 kVp/40 mAs, 70 kVp/80 mAs and 80 kVp/40 mAs for the 1-, 5- and 12-year-old-equivalent phantoms respectively (kappa = 0,70). Optimised dose settings reduced CTDI(w) doses 89-93%. The effective dose was 0.5 mSv (91-94,5% reduction). The optimised protocols were used clinically without problems.

Conclusions: Radiation doses for intraoperative 3-D CT using a cone-beam flat-detector scanner could be reduced at least 89% compared to manufacturer settings and still be used to safely navigate pedicle screws.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3414709PMC
http://dx.doi.org/10.1007/s00247-012-2396-0DOI Listing

Publication Analysis

Top Keywords

cone-beam flat-detector
12
intraoperative cone-beam
8
pedicle screws
8
ctdiw doses
8
kvp/40 mas
8
dose
4
dose optimisation
4
intraoperative
4
optimisation intraoperative
4
flat-detector paediatric
4

Similar Publications

Background: In spite of the tremendous potential of game-changing biological image- and/or biologically guided radiation therapy (RT) and adaptive radiation therapy for cancer treatment, existing limited strategies for integrating molecular imaging and/or biological information with RT have impeded the translation of preclinical research findings to clinical applications. Additionally, there is an urgent need for a highly integrated small-animal radiation therapy (SART) platform that can seamlessly combine therapeutic and diagnostic capabilities to comprehensively enhance RT for cancer treatment.

Purpose: We investigated a highly integrated quad-modal on-board imaging configuration combining positron emission tomography (PET), single-photon emission computed tomography (SPECT), photon-counting spectral CT, and cone-beam computed tomography (CBCT) in a SART platform using a Monte Carlo model as a proof-of-concept.

View Article and Find Full Text PDF

Pre-operative embolization of hypervascular intracranial tumors can be performed to reduce bleeding complications during resection. Accurate vascular mapping of the tumor is necessary for both the correct indication setting for embolization and for the evaluation of the performed embolization. We prospectively examined the role of whole brain and selective parenchymal blood volume (PBV) flat detector computer tomography perfusion (FD CTP) imaging in pre-operative angiographic mapping and embolization of patients with hypervascular intracranial tumors.

View Article and Find Full Text PDF

Background: Flat detector computed tomography (FDCT) is widely used for periprocedural imaging in the angiography suite. Sine Spin FDCT (SFDCT) is the latest generation of cone beam CT using a double oblique trajectory for image acquisition to reduce artefacts and improve soft tissue brain imaging. This study compared the effective dose, image quality and diagnostic performance of the latest generation of SFDCT with multidetector CT (MDCT).

View Article and Find Full Text PDF

Purpose: Conventional cone-beam computed tomography CT (CBCT) provides limited discrimination between low-contrast tissues. Furthermore, it is limited to full-spectrum energy integration. A dual-energy CBCT system could be used to separate photon energy spectra with the potential to increase the visibility of clinically relevant features and acquire additional information relevant in a multitude of clinical imaging applications.

View Article and Find Full Text PDF

The technical feasibility and diagnostic potential of angiographic flat-detector perfusion imaging technique, combining digital subtraction angiography with a flat-detector computed tomography steady-state perfusion imaging, was explored in patients treated with direct or indirect revascularization surgery. This short communication is about an imaging modality with great potential for evaluation, comparison and grading of vascular perfusion territory areas and anatomical location selectively perfused by direct and indirect cerebral bypasses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!