Objectives: The purpose of this study was to demonstrate the feasibility and procedural outcomes with a new self-expanding and repositionable transcatheter heart valve.
Background: Transcatheter aortic valve replacement is a viable option for selected patients with severe symptomatic aortic stenosis. However, suboptimal prosthesis positioning may contribute to paravalvular regurgitation, atrioventricular conduction block, and mitral or coronary compromise.
Methods: The repositionable Portico valve (St. Jude Medical, Minneapolis, Minnesota) was implanted in 10 patients with severe aortic stenosis utilizing percutaneous femoral arterial access. Patients underwent transthoracic and transesophageal echocardiography and multidetector computed tomography before and after valve implantation. Clinical and echocardiographic follow-up was obtained at 30 days.
Results: Device implantation was successful in all patients. Prosthesis recapture and repositioning was performed in 4 patients. Intermittent prosthetic leaflet dysfunction in 1 patient required implantation of a second transcatheter valve. There was 1 minor stroke. At 30-day follow-up, echocardiographic mean transaortic gradient was reduced from 44.9 ± 16.7 mm Hg to 10.9 ± 3.8 mm Hg (p < 0.001), and valve area increased from 0.6 ± 0.1 cm(2) to 1.3 ± 0.2 cm(2) (p < 0.001). Paravalvular regurgitation was mild or less in 9 patients (90%) and moderate in 1 patient (10%). There were no major strokes, major vascular complications, major bleeds, or deaths. No patient required pacemaker implantation. All patients were in New York Heart Association functional class II or less.
Conclusions: Transcatheter aortic valve replacement with the repositionable Portico transcatheter heart valve is feasible, with good short-term clinical and hemodynamic outcomes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacc.2012.02.045 | DOI Listing |
The authors propose a modified transcatheter aortic valve replacement technique wherein the removal of the guidewire and delivery catheter immediately after the valve implantation helps to not only shorten the procedure but also decrease complications induced by the guidewire and delivery catheter.
View Article and Find Full Text PDFEchocardiography
January 2025
Cardiology Department, Soroka University Medical Center, Beer-Sheba, Israel.
Background: Timing of treatment of aortic stenosis (AS) is of key importance. AS severity is currently determined by transthoracic echocardiography (TTE) with a main focus on mean trans-aortic gradients. However, echocardiography has its limitations.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
January 2025
Department of Cardiovascular Medicine, Osaka University Graduate School of Medicine, Osaka, Japan.
Background: The long-term valve durability of supra-annular self-expanding valves (SEV) and intra-annular balloon-expandable valves (BEV) in patients with small aortic annuli remains unexplored.
Aims: This study aimed to determine the long-term bioprosthetic valve durability with SEV versus BEV in patients with small aortic annuli.
Methods: This retrospective study included patients with severe aortic stenosis (AS) and an aortic annulus area of 430 mm or less who underwent transcatheter aortic valve replacement using SEV and BEV between October 2009 and December 2022.
EuroIntervention
January 2025
Department of Cardiology, Universitair Ziekenhuis Brussel, Vrije Universiteit Brussel, Jette, Belgium.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!